Safety Precautions
To ensure safe operation, follow the precautions below.
When a gas alarm activates, immediately take all the necessary
measures to prevent an explosion.
Air adjustment (zeroing) starts automatically when the detector is
turned on. Make sure to turn on the detector in clean air. Failure
to do so may cause incorrect air adjustment which will then lead
to inaccurate measurement being displayed.
This product is safety equipment. Perform a routine check before
use (page 35 “Routine Check”).
Do not cover any gas detection ports with any fabric, etc. If
blocked, correct detection is not possible.
Do not block the speaker opening. If blocked, the audible alarm
will be muffled.
Keep the filter elements at the gas detection ports clean and dry.
If the filter elements are dirty or wet, proper gas detection is not
possible.
The recommended sensor replacement cycle is two years. The
sensor may fail to provide accurate detection after two years, and
should be replaced.
Use the specified battery. Use of batteries other than specified
may impair the product’s intrinsically safe characteristics (page 28
“Battery Replacement”).
WARNING
DANGER
- 3 -
Explosion-proof Requirements
Follow the conditions below to comply with the explosion-proof requirements.
Symbols Used in this Manual
This manual uses Danger, Warning, Caution and Note symbols to draw attention to
procedures, materials, methods, and processes, which require particular attention.
Indicates an imminently hazardous situation that can result in
death or serious injury.
Indicates a potentially hazardous situation that may result in
death or serious injury.
Indicates a hazardous situation that may result in minor injury
or property damage.
Provides information on product handling.
ATEX Explosion-proof: II 1 G Ex ia IIC T3 Ga
Certificate No.: CML 15ATEX2117X
Standard:
EN 60079-0:2018
EN 60079-11:2012
IECEx Explosion-proof: Ex ia IIC T3 Ga
Certificate No.: IECEx CML 17.0009X
Standard:
IEC 60079-0:2017 Ed.7
IEC 60079-11:2011 Ed.6
Rating: 1.5_VDC (LR03 manufactured by Panasonic x 1)
1.5_VDC (LR03 manufactured by Duracell x 1)
1.5_VDC (LR03 manufactured by Energizer x 1)
1.5_VDC (LR03 manufactured by Varta x 1)
1.3_VDC (GP75AAAHC manufactured by GP x 1)
Ambient temperature: -20 to 50
C
Conditions of use:
・
Replace the battery in a non-hazardous area.
・
Charge the battery in a non-hazardous area.
・
Charge the battery with a specified battery charger (BC-9).
・
Do not separate XA-4400ll and BP-4000llAL or BP-4000llMH when an explosive
gas atmosphere may be present.
・
To prevent accidents from electrostatic charges, wear anti-static working clothes
and conductive footwear (antistatic working shoes). Additionally the work floor
should be conductive (leak current: 10M ohm or less).
・
When the Energizer Type E92 AAA LR03 is used, the ambient temperature
range is -18 to +50
C.
・
The enclosure of this detector uses ABS resin. Avoid exposure to organic
solvents, strong acids, strong alkalis and mineral oils. Exposure may
compromise the explosion-proof structure of the detector.
WARNING
NOTE
DANGER
WARNING
CAUTION