5 Screen List
5
Timer
Pressurization (CHG)
0 to 999.9 [s]
Equalization (BAL1)
Stabilization (BAL2)
Detection (DET)
Test
Pressure
Pressure Unit
kPa, MPa (PSI, kg/cm
2
, bar, mbar, mmHg, cmHg, inHg)
*1
Pressure Setting
Pressure applied to tested part (WORK) and reference part (MASTER)
Upper Press Limit
Varies depending on the test pressure range and unit.
Lower Press Limit
Leak Limit
Leak Unit
Pa, kPa, mL/s, mL/min, L/min, Pa·m
3
/s, E-3 Pa·m
3
/s, Pa/s, Pa/min,
*Pa/s, *Pa/min (mmH
2
O, inH
2
O, mmHg, in
3
/min, in
3
/d, ft
3
/h)
*1
DET UL
Small leak limit for WORK side during DET stage
DET LL
Small leak limit for MASTER side during DET stage
K(Ve)
K(Ve) Unit
Unit for K(Ve) value
K(Ve) Value
Enter K(Ve) value if known.
*1
The units in ( ) are not available for SI unit restriction models.
3.2 Advanced Setting
All the leak test settings including the basic settings
Advanced Settings
Timer
Test Press
Title
Unit
Channel Title
Leak Unit
Pressure Unit
K(Ve) Unit
Charge Delay (DL1)
PCHK
Pressurization (CHG)
Equalization (BAL1)
Balance Delay (DL2)
Stabilization (BAL2)
Detection (DET)
Air-Blow (BLW)
Exhaust (EXH)
End (END)
Pressure Setting
Upper Press Limit
Lower Press Limit
Pressure Monitoring
PS Auto-Zero