Error message:

ERROR XX is displayed, and the power supply will stop when the current reaches the limit values.

Code 00:

Current is too low.

Code 01:

Current is too high.

4. Sample Loading 

Gently load the samples into the well using a micropipette. Add small volumes of glycerol or

sucrose to the sample facilitate sample. The maximum loading volume of the f sample is about 18

µ

R

for the 13 wells comb and 9 µ

R

for the 26 wells comb for the thick gel. If you use a

multichannel pipette, use the loading guide plate and follow one of the proposed patterns:

5. Migration 

Place the cover on the electrophoresis tank. This will close the electrical circuit.

If the buffer temperature rises above 40°C during electrophoresis, turn off the power and either

dilute the running buffer or place i-MyRun in a cold room. Be aware that if the buffer

concentration is too high, the temperature will increase rapidly.

If you want to withdraw the agarose gel from the support, first cut the two small overhangs of

agarose located at the end of the gel.

Note: Refer to page 4.

6. Electrical Specification

Voltage from 100 to 240 VA, AC 50/60 Hz

Temperature of operation: from 0–40°C

Output constant voltage: 35, 50, 75, 100, 135

VDA Max 54 W or 400 mA

Output constant ampere: 20, 30, 40, 50, 60, 70, 80 mA max 24 W or 300 VG

5. Maintenance

Wipe the instrument regularly with a damp cloth. Let the instrument dry completely before use.

Otherwise, no user maintenance is necessary. All servicing should be entrusted to qualified

personnel 

only. Please contact Cosmo Bio Co., Ltd., for more service information.

6. Environmental Condition

1. _Indoor use.

2. _Altitude up to 2000 m.

3. _Temperature 5°C to 40°C.

4. _Maximum relative humidity of 80% for temperature up to 31°C decreasing to 50% relative 

humidity at 40°C.

5. _Pollution Degree 2 in accordance with IEC 664.

7

Warranty; One year

MyRunN/manual  04.10.28  10:48 AM  ページ9

Содержание i-MyRun.N

Страница 1: ...SMO BIO CO LTD Toyo Ekimae Bldg 2 20 Toyo 2 Chome Koto ku Tokyo 135 0016 Japan Tel 81 3 5632 9605 Fax 81 3 5632 9614 i MyRun N Instruction Manual Electrophoresis System MyRunN manual 04 10 28 10 48 AM...

Страница 2: ...IMR 201 Electrophorsis Tank for i MyRun N IMR 301 Gel Casting System for i MyRun N IMR 401 Support guide for Minigel L S IMR 501 Terms and Conditions of Sale All goods and services are sold subject t...

Страница 3: ...sed in the operation of this instrument as it can develop sufficient voltage and current to produce a lethal shock To avoid any risk of injury the instrument should only be operated by properly traine...

Страница 4: ...ld the power supply be connected to an electrophoresis unit that has a leakage path to earth the IMR 201 will detect this fault and the high voltage will be turned off 2 _Start current check To ensure...

Страница 5: ...e for use on six lanes A1 A6 1 _Make the gels by using the 26 tooth comb on six lanes A1 A6 2 _Set a gel with gel casting plate on the electrophoresis tank 3 _Set the guide over the plate 4 _Load the...

Страница 6: ...art of the gel unit with organic solvents such as ethanol or acetone 3 Power Supply Operation Check the Quick Start Operation Manual 1 _Insert the power supply connector into the electrophoresis tank...

Страница 7: ..._The display shows the last value of the selected and keys 3 _Select and adjust the running time by the selecting the key followed by 4 _Select If you do not wish to use the timer 1 _After selecting t...

Страница 8: ...centration is too high the temperature will increase rapidly If you want to withdraw the agarose gel from the support first cut the two small overhangs of agarose located at the end of the gel Note Re...

Отзывы: