3.1.3. Opening of shut-off valves
Open the shut-off valve with identification plate "instrument air supply".
3.1.4. Setting of gas pressure reducer
Turning the adjusting cap clockwise can raise the pressure of the gas pressure reducer.
The set outlet pressure can be read immediately from the pressure indicator mounted on
the reducer.
Set the output gas pressure to 30 PSIG.
3.1.5. Adjusting flow with flow meters
Unscrew the needle valve of bypass flow regulator completely (turn counter clockwise).
Use the "By-pass flow meter" needle valve; throttle back the flow so that the "Analyzer
flow meter" shows a flow of 50Nl/hr.
After this throttle back the output of the analyzer flow to 40Nl/hr using the needle valve
on flow meter tagged "Analyzer flow".
3.1.6. Adjusting booster relays
Set the chosen residual O
2
set point with help of the booster relay. With the booster
relay it is possible to give a small negative (i.e. the gas pressure is higher than the air
pressure) or positive offset of max 0.5 bar, as set point correction or as pressure drop
compensation over the air- and gas tubes in the heat exchanger.
In order to achieve the necessary offset pressure, the insert with internal hexagon at the
top of the booster relay must be removed. By doing this, the slotted adjusting screw is
accessible by means of a small rotation counter clockwise, a bigger negative offset is
reached. In other words, the air pressure is lowered in comparison with the gas
pressure. The correct choice of a negative rather than a positive offset needs to be used
for proper analyzer operation. The chosen residual oxygen concentration, combined with
the average of the chosen range of measurement, gives according to these curves the
selection of the air nozzle and an indication for the booster offset.
Содержание 9610
Страница 27: ...INSTALLATION OPERATION AND MAINTENANCE MANUAL COSA 9610 Page 26 of 48 ...
Страница 32: ...INSTALLATION OPERATION AND MAINTENANCE MANUAL COSA 9610 Page 31 of 48 3 3 2 Operation Menu ...
Страница 34: ...INSTALLATION OPERATION AND MAINTENANCE MANUAL COSA 9610 Page 33 of 48 3 3 4 Output Menu ...
Страница 36: ...INSTALLATION OPERATION AND MAINTENANCE MANUAL COSA 9610 Page 35 of 48 ...
Страница 38: ...INSTALLATION OPERATION AND MAINTENANCE MANUAL COSA 9610 Page 37 of 48 Display Menu ...
Страница 39: ...INSTALLATION OPERATION AND MAINTENANCE MANUAL COSA 9610 Page 38 of 48 3 3 7 Reset Alarms Menu ...
Страница 46: ...INSTALLATION OPERATION AND MAINTENANCE MANUAL COSA 9610 Page 45 of 48 5 INSTALLATION DRAWING ...
Страница 48: ...INSTALLATION OPERATION AND MAINTENANCE MANUAL COSA 9610 Page 46 of 48 ...
Страница 50: ...INSTALLATION OPERATION AND MAINTENANCE MANUAL COSA 9610 Page 48 of 48 ...
Страница 51: ...INSTALLATION OPERATION AND MAINTENANCE MANUAL COSA 9610 Page 48 of 48 ...