MOBILTEX® DATA LTD.
Calgary, Alberta
www.corTalk.com
TITLE:
uDL1 Configuration and Operation Guide
DOCUMENT NO.:
UDL1-MAN-001
SHEET:
14 of 26
REV:
1.07
Figure 14 Detailed Status Dialog
4.3.1.11
Reboot uDL1
The “Reboot uDL1” is used to restart the uDL1 application.
4.3.2
Configuration Group
All items in the configuration group may be reset back to factory defaults by clicking on the “Defaults” button.
4.3.2.1 Timing
The “Timing” section allows you to select the measurement sampling interval (1 second to 12 hours).
An additional parameter “1s DC Sampling Delay” can be used when DC acquisition is enabled and the sampling interval
is 1 second. This parameter delays the acquisition of a data sample from the top of the second. This could be used when
a signal is expected to transition co-incident with the transition of the second, but there is an associated settling time
with the signal transition. Delaying the sampling past the settling time will then yield a post-transition reading. This
parameter is only useful on advanced uDL1 models. Normally, when not needed, the parameter should be left at the
default value of 100ms to minimize battery consumption during unit operation.
4.3.2.2 Delayed Start Timing (v1.16+)
Starting with version 1.16 firmware, the uDL1 can now be programmed to delay data acquisition start by a programmable
number of minutes. This allows the unit hardware to be deployed days or weeks in advance of a desired measurement
period. The acquisition start delay can be set to a value between 0 and 40320 minutes (28 days). When the analog cable
is plugged into the unit, it will perform its normal first minute of measurement previews and then enter the delayed start