Guidance and manufacturer’s
declaration
—
electromagnetic
immunity
Corsano bracelet is intended for use in the electromagnetic environment specified below. The customer or the
user of Corsano bracelet should assure that it is used in such an environment.
Immunity test
IEC 60601
test
level
Compliance
level
Electromagnetic environment - guidance
Conducted
3 Vrms
NA
Portable and mobile RF communications equipment
should be used no closer to any part of Corsano bracelet,
including cables, than the recommended separation
distance calculated from the equation applicable to the
frequency of the transmitter.
Recommended separation distance (m)
d = 1.2
√
P
d = 1.2
√
P 80 MHz to 800 MHz
d = 2.3
√
P 800 MHz to 2.5 GHz
where
P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and
d is the recommended separation
distance in meters (m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey
a
,
should be less than the compliance level in each
frequency range
b
.
Interference may occur in the vicinity of equipment
marked with the following symbol:
Conducted RF
IEC 61000-4-6
150 kHz to
80
MHz
NA
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz to
2.5 GHz
NA
NOTE 1
—
At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2
—
These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects, and people
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land
mobile radios, amateur radio, AM and FM radio broadcast, and TV broadcast cannot be predicted theoretically
with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic
site survey should be considered. If the measured field strength in the location in which Corsano bracelet is
used exceeds the applicable RF compliance level above, Corsano bracelet should be observed to verify normal
operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting
or relocating Corsano Bracelet.
Содержание CardioWatch 287
Страница 1: ...As of 2021 11 23 Revision status 9 1 EN CardioWatch 287 1B Bracelet Instruction Manual...
Страница 17: ...As of 2021 11 23 Revision status 9 17 After you have signed up you can Sign In...
Страница 24: ...As of 2021 11 23 Revision status 9 24 Sleep...
Страница 25: ...As of 2021 11 23 Revision status 9 25 Heart Rate...
Страница 35: ...As of 2021 11 23 Revision status 9 35...
Страница 49: ......