12 CLINICAL PERFORMANCE
Definition
Unit
Range
Acquisition
Time
Update
interval
Accuracy
*
Heart Rate
Number of beats of the heart
per minute
beats/minute
(bpm)
30-220
5-10 sec
1 sec
HR is ±4 bpm MAD,
±5% MARD (at rest)
RR Interval
Elapsed time between two
consecutive heart beats
msec
300-2000 ms 5-10 sec
1 sec
RR Interval ±50 ms
MAD, ±5% MARD (at
rest)
Heart Rate
Variability
Beat to beat (RR interval)
variations
msec
0-200 ms
5-10 sec
1 sec
HRV ±10 ms MAD,
±5% MARD (at rest)
Respiration
Rate
Number of breaths
(inhalation - exhalation
cycles) per minute
breaths/minute
(brpm)
5-45 brpm
20-30 sec
1 sec
Respiration Rate ±1
bprm MAD, ±5%
MARD (at rest)
Sleep
Stages
Detection of specific sleep
stages & sleep HR
awake, light
sleep, deep
sleep, REM
sleep stage
upon end of
the entire
sleep event
1 min
Sleep Stage ±10 %
MAD
Sleep Score
Sleep performance and sleep
consistency with equal
weight
%
0-100%
10 sec
1 sec
Sleep Score ±5 %
MAD
Remarks:
* MAD=Mean absolute difference & MARD=Mean absolute relative difference under motion
Содержание CardioWatch 287
Страница 1: ...As of 2021 11 23 Revision status 9 1 EN CardioWatch 287 1B Bracelet Instruction Manual...
Страница 17: ...As of 2021 11 23 Revision status 9 17 After you have signed up you can Sign In...
Страница 24: ...As of 2021 11 23 Revision status 9 24 Sleep...
Страница 25: ...As of 2021 11 23 Revision status 9 25 Heart Rate...
Страница 35: ...As of 2021 11 23 Revision status 9 35...
Страница 49: ......