3.3 Procedure to Adjust Scaling for Engineering Units 

Step 1 

Apply power and allow the amplifier to warm up for 10-15 minutes. 

 
Step 2 

With zero load or pressure on the transducer, adjust the ZERO potentiometers to indicate the 

desired zero on the readout instrument. 

 
Step 3 

Connect a jumper across PINS 5 & 6 of the power terminals (See Figure 1-1) or wire in a switch 

as shown in the wiring section. Adjust the SPAN potentiometer to the calculated voltage or calculated 
units on the readout instrument. 

 
Step 4 

Disconnect the shunt cal jumper or release the shunt cal switch and repeat steps 2 

& 3 if needed. 

 

Warranty 

Limited Warranty on Products  

Any Cooper Instruments product which, under normal operating conditions, proves defective in material or in 
workmanship within one year of the date of shipment by Cooper will be repaired or replaced free of charge 
provided that a return material authorization is obtained from Cooper and the defective product is sent, 
transportation charges prepaid, with notice of the defect, and it is established that the product has been properly 
installed, maintained, and operated within the limits of rated and normal usage.   

Replacement or repaired product will be shipped F.O.B. from our plant.  The terms of this warranty do not extend 
to any product or part thereof which, under normal usage, has an inherently shorter useful life than one year.  
The replacement warranty detailed here is the buyer’s exclusive remedy, and will satisfy all obligations of Cooper 
whether based on contract, negligence, or otherwise.  Cooper is not responsible for any incidental or 
consequential loss or damage which might result from a failure of any and all other warranties, express or implied, 
including implied warranty of merchantability or fitness for particular purpose.  Any unauthorized disassembly or 
attempt to repair voids this warranty. 

Obtaining Service under Warranty 

Advance authorization is 

required

 prior to the return to Cooper Instruments.  Before returning the item, contact 

the Repair Department c/o Cooper Instruments at (540) 349-4746 for a Return Material Authorization number.  
Shipment to Cooper shall be at buyer’s expense and repaired or replacement items will be shipped F.O.B. from 
our plant in Warrenton, Virginia.  Non-verified problems or defects may be subject to a $150 evaluation charge.  
Please return the original calibration data with the unit. 

Repair Warranty 

All repairs of Cooper products are warranted for a period of 90 days from date of shipment.  This warranty applies 
only to those items that were found defective and repaired; it does not apply to products in which no defect was 
found and returned as is or merely recalibrated.  It may be possible for out-of-warranty products to be returned 
to the exact original specifications or dimensions. 

* Technical description of the defect: In order to properly repair a product, it is 

absolutely necessary

 for Cooper to 

receive information specifying the reason the product is being returned.  Specific test data, written observations 
on the failure and the specific corrective action you require are needed.  

 