5

Operation

The Sperm Preparation Reader detects tagged containers that are brought into the work area. 

When a sample is to be transferred between containers, bring both containers into the work 

area. The Sperm Preparation Reader will detect the tags and the RI Witness system will record 

the step.

Cleaning 

The Sperm Preparation Reader may be cleaned with a soft cloth and mild detergent. The device 

may be lifted and returned to its original location. Do not disconnect the cables attached to the 

device.

SECTION 6 - TROUBLESHOOTING

RFID System

Problem

Possible Cause

Solution

Tags Not Reading

Metal near reader

Remove any metallic objects from the area, 

check if the tags reappear in the Work Area

Loose or no 

connection

Check security of USB and power cable 

connections. Verify that the light on the power 

supply is illuminated

RF noise or 

interference

Other electrical devices in the lab can cause 

RF noise/interference. If a portable electronic 

device has been brought close to the device, 

remove it and check if the tags reappear in 

the Work Area

Broken tag

Check whether the tag is readable by a 

different RI Witness device. If it is not, discard 

that tag

Antenna tuning 

problem

Navigate to the 

Work Area Settings

 screen, 

then click 

Connected Devices

, then 

Sperm 

Prep Reader

, then 

RFID Tuning

, check that 

all 3 channels have green ticks next to them. 

If any have a yellow warning triangle next to 

them, contact an RI service representative

Содержание RI Witness 6-70-854

Страница 1: ...sales coopersurgical com fertility coopersurgical com Research Instruments Ltd Bickland Industrial Park Falmouth Cornwall TR11 4TA UK Document 6 70 854UM 5 DRF 5438 9th June 2020 RI Witness Sperm Pre...

Страница 2: ...information If you are unsure of any of the information contained in this manual then you should contact Research Instruments or an appointed representative before attempting to use this equipment In...

Страница 3: ...tion 6 70 854 RI Witness Sperm Preparation Reader SECTION 3 SAFETY WARNINGS DO NOT disassemble or modify any part of the RI Witness Sperm Preparation Reader or substitute any component for any other D...

Страница 4: ...pment is operated in a commercial environment This equipment generates uses and can radiate radio frequency energy and if not installed and used in accordance with the instruction manual may cause har...

Страница 5: ...the footprint of the product Power Supply Input 100 240VAC 50 60Hz 3A Class I 0 8 0 4A Output 12VDC 2 5A Max 30W USB USB 2 0 Socket Type Mini B For connection to tablet or PC approved to IEC 60950 1...

Страница 6: ...Cause Solution Tags Not Reading Metal near reader Remove any metallic objects from the area check if the tags reappear in the Work Area Loose or no connection Check security of USB and power cable co...

Страница 7: ...ines that the product fails to conform to that warranty during the one year period RI will repair or replace the product at RI s discretion free of charge To return the product to RI a customer must c...

Страница 8: ...ice Please provide customer service with full details of the incident including any applicable serial numbers In some instances it may be necessary to return the device to the manufacturer to assist i...

Страница 9: ......

Отзывы: