NOTES: 

Содержание EnviroTrak Sensorex TX100

Страница 1: ...pH Monitoring Installation and Operating Guide Cooper Atkins Corporation Accuracy to the Highest Degree www cooper atkins com ...

Страница 2: ...2 ...

Страница 3: ...oper Atkins Please feel free to contact I Care at 888 533 6900 x2 for assistance Package Contents 3 Hardware Mounting 4 A Mounting 4 pH Electrode Calibration 5 A Calibrating pH Electrode 7 00pH and 4 00pH 5 pH Electrode Clean Electrode Notification 6 A Setting Cleaning Reminder 6 Transmitter Setup 7 Software Configuration pH Electrode 8 ...

Страница 4: ...ex pH Monitor D Analog Transmitter a Transmitter Box b Analog Transmitter wired to Sensorex pH Monitor E pH Electrode and Cable F pH Power Plug Note 4 01pH and 7 0pH calibration solutions NOT INCLUDED but required for calibrating pH electrode Needs to be purchased seperately Not sold by Cooper Atkins A F B Da Db C E ...

Страница 5: ...of mounting options Please see mounting directions in the Sensorex Installation and Operating Manual 4 Typically the Sensorex pH Monitor box should be mounted at eye level 5ft to 6ft for easy reading SAMPLE ELECTRODE MOUNTING MONITORING WATER LINE FOR ADDITIONAL MOUNTING INSTRUCTIONS REFER TO THE SENSOREX MANUAL Note Installation of probe is the responsibility of the facility Please consult local ...

Страница 6: ...will show the expected buffer 7 00 or 6 86 for the first point Use the UP and DOWN arrows to select 7 00 Place the electrode into the 7 00 buffer solution Press and hold E to start 1 st Calibration 4 When the reading is stable Press E to save first point 5 Now the small display will show the 2 nd expected buffer 4 01 9 18 or 10 00 Use the UP and DOWN arrows to select 4 01 Place the electrode into ...

Страница 7: ...EX MANUAL PAGE 25 IF NEEDED A SETTING CLEANING REMINDER 1 Press and hold E to enter setup 2 Press UP arrow 2 times C P will display on the main display and Clean and electrode icon will flash 3 Press E and OFF will flash 4 Use the UP and DOWN arrows to turn ON Clean Probe 5 Press E to accept Day number will flash The default value is 90 6 Use the UP and DOWN arrows to change number of days 7 Press...

Страница 8: ...newly configured pH values 900Mhz Digital Analog Transmitter 1 Remove the Transmitter cover 2 Verify that the Transmitter is wired D for 4 20mA input 3 Verify that the Transmitter Jumper C is configured for 4 20mA input 4 Insert the battery s into the Battery Slot A 5 Press the Reset Button B to initialize the Transmitter 6 Install the Transmitter cover D C ...

Страница 9: ...g point 1 Log into EnviroTrak 2 To register a pH Electrode click the Configuration EnviroTrak Device Registration Sensor Registration 3 The sensor registration information is located inside the lid or on the supplied stickers 4 Enter the Sensor ID A Security ID B select DIGITAL RF Analog Input C as the transmitter type and then select Sensorex TX100 pH Monitor D as the 1 input Note C Make sure you...

Страница 10: ...ransmitter to from the Assign To Group Optional E drop down list Note If no group is selected the sensor will only display for users using the Default All Sensors view 6 Click Add F You will be asked if you wish to name the transmitter Follow the directions and rename the transmitter or close the Set Sensors window You should see the following confirmation at the bottom of the Transmitter Registra...

Страница 11: ...11 NOTES ...