          Version Number: 1.0 

 

4. EMC Declaration      (Continued)                                                                                       

Guidance and manufacturer’s declaration – electromagnetic emissions

 

The “Digital Blood Pressure Monitor B61T” is intended for use in the electromagnetic environment specified below. The 
customer or the user of the “Digital Blood Pressure Monitor B61T” should ensure that it is used in such an environment.

 

Emissions test 

Compliance 

Electromagnetic environment 

– 

guidance 

RF emissions 

CISPR 11

 

Group 1

 

The “Digital Blood Pressure Monitor B61T” uses RF energy only for its 
internal function. Therefore, its RF emissions are very 
low and are not likely to cause any interference in   
nearby electronic equipment.

 

RF emissions 

CISPR 11

 

Class B

 

The “Digital Blood Pressure Monitor B61T” is suitable for use in all 
establishments, including domestic establishments and 
those directly connected to the public low-voltage 
power supply network that supplies buildings used for 

domestic purposes.

 

Harmonic emissions 

IEC 61000-3-2

 

Class A

 

 

Voltage fluctuations/ 
flicker emissions 

IEC 61000-3-3

 

Complies 

 

 

 

Recommended separation distances between 

portable and mobile RF communications equipment and the Medical Digital Blood 

Pressure Monitor B61T 

The “Digital Blood Pressure Monitor B61T” is intended for use in an electromagnetic environment in which radiated RF 
disturbances 
are controlled. The customer or the user of the Medical Digital Blood Pressure Monitor B61T can help prevent 
electromagnetic interference 
by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the 
“Digital  Blood  Pressure  Monitor  B61T”  as  recommended  below,  according  to  the  maximum  output  power  of  the 
communications 
equipment.

 

Rated maximum output 

power of transmitter 

W

 

Separation distance according to frequency of transmitter 

m

 

 

150 kHz to 80 MHz 

 

 

80 MHz to 800 MHz 

 

 

800 MHz to 2,5 GHz 

 
 
 

0,01 

0.12 

0.12 

0.23 

0,1 

0.38 

0.38 

0.73 

1.2 

1.2 

2.3 

10 

3.8 

3.8 

7.3 

100

 

12 

12 

23 

For transmitters rated at a maximum output power not listed above, the recommended separation distance 

in 

meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where 

is the 

maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer. 
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.  
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and 
reflection from structures, objects and people.

 

 

 

 

 

Содержание B61T

Страница 1: ...ill prove helpful in allowing you to enjoy your measurement in hospitals and physician s offices throughout the world where accuracy and dependability are critical professional diagnostic products are...

Страница 2: ...using this device Blood pressure readings obtained by this device should be verified before prescribing or making adjustments to any medications used to control hypertension Under no circumstances sh...

Страница 3: ...e Warning This equipment shall not be serviced or maintained while in use with the patient Warning The patient is the intended operator the functions of monitoring blood pressure and pulse rate can be...

Страница 4: ...mHg while at rest In this case please consult your physician immediately Long term values at this level endanger your health due to continual damage to the blood vessels in your body If your systolic...

Страница 5: ...lerosis Furthermore the heart will become structurally damaged with increased blood pressure values 3 There are many different causes of high blood pressure We differentiate between the common primary...

Страница 6: ...high blood pressure 4 2 System Settings After the battery is fully recharged some functions can be set before use this monitor a Start Stop Press the Start Stop button one time The indicator around th...

Страница 7: ...measurement error of 10 mmHg Cuffs that are too narrow or too short result in false measurement values Selecting the correct cuff is extremely important Cuff size is dependent upon the circumference...

Страница 8: ...ed on the arm and connected to the blood pressure monitor the measurement can begin a Operate on the device Press the Stop Start button as the power on blue indictor is turned on The pump begins to in...

Страница 9: ...After the monitor has detected your blood pressure and pulse rate the cuff automatically deflates Your blood pressure will be voiced out or shown on the app on smart phone if operated by App 7 The mo...

Страница 10: ...ted with the Digital Blood Pressure Monitor B61T and its use Authorized Representative in the European Community CE Mark conforms to essential requirements of the Medical Device Directive 93 42 EEC Da...

Страница 11: ...sed by different patients wash hands before and after each use a Do not expose the device to either extreme temperatures humidity dust or direct sunlight b The cuff contains a sensitive air tight bubb...

Страница 12: ...e And Shipping Conditions Temperature 25 to 70 Humidity 10 to 93 RH Pressure altitude 70KPa 106Kpa Measuring method Oscillometric Pressure sensor Capacitive Measuring range DIA 40 130mmHg SYS 60 230mm...

Страница 13: ...ast 30 Electrical fast transient burst IEC 61000 4 4 2 kV for power supply lines 1 kV for input output lines 2 kV for power supply lines 1 kV for input output lines Mains power quality should be that...

Страница 14: ...netic field IEC 61000 4 8 3 A m 3 A m Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment NOTE UT is the a c mains...

Страница 15: ...om fixed RF transmitters as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked...

Страница 16: ...d Pressure Monitor B61T The Digital Blood Pressure Monitor B61T is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled The customer or the user of the M...

Страница 17: ...Version Number 1 0 10 Warranty Card...

Страница 18: ...n Jamr Medical Technology Co Ltd Address 2nd Floor A building No 2 Guiyuan Road Guihua Community Guanlan Town Longhua New District Shenzhen China Authorized Indian Representative Cooey Technologies Pv...

Отзывы: