BOOM & SPREADER BAR

Check connections between Boom and Mast for 
improper connection, looseness or wear.

Check the Boom for bending and deflection.

MAST

Check the Mast for bending or deflection.

BASE & FOOT PEDAL

Check bolts and nuts for looseness..

Check casters and axle bolts for tightness.

Check rubber parts on the casters for deflection

20

MAINTENANCE & INSPECTION

The operator of the lift shall inspect the ConvaQuip Lift before each 
use. Check all bolts for tightness. Make sure the base can be easily 
widened and that all lift parts are in place. Make sure that casters can 
be turned freely, and that caster brakes can be engaged. Make certain 
all necessary items (i.e. slings, wheelchairs, etc.) are ready for use.

At least once a month, the lift should be thoroughly inspected by 
a person qualified to recognize any signs of wear and tear and 
looseness of bolts or parts. Replace any worn parts immediately.

To lubricate, put a drop of oil on the following points when the 
ConvaQuip Lift is placed into service and every month thereafter: Top 
of Mast, Spreader Hinge, and Caster Axles.

FIRST

RECEIVED

MONTHLY

EVERY 3 

MONTHS

BASE & FOOT PEDAL

Apply grease to caster ball bearings if 
needed.

Check welding joints for rust and crack.

CLEANING

 

AS NEEDED

ACTUATOR & CONTROL BOX

Make sure the control box is firmly affixed 
to the mast.

Make sure the actuator is secured to the 
Boom and mast with pins and key rings

Make sure actuator plug into control box is 
not loose.

SLINGS & SLING HARDWARE

Check sling for wear.

Check sling hardware every time before 
use.

21

Содержание STS600E

Страница 1: ...onvaQuip STS600E Ensure the product has been assembled according to the instructions in this manual All operators should have received or read and understood the instructions for safe and proper opera...

Страница 2: ...must be taken with patients who are comatose spastic agitated or otherwise severely handicapped A competent caregiver must always be present during any patient lifting The sole function of the patien...

Страница 3: ...and components from other manufacturers Service and repair of the patient lift should be performed by ConvaQuip or its authorized dealers ONLY ConvaQuip is not responsible for any consequence resultin...

Страница 4: ...pport Strap Quick Release Dual Front Casters 600 Lb 68 33 25 5 38 4 5 40 19 Yes Optional Optional Optional Yes Yes PRODUCT FEATURES ConvaQuip is committed to continuous improvements of our products th...

Страница 5: ...tor to the bracket on the Mast and insert pin Insert keeper ring through hole in pin to secure base of actuator B Attach the top of the Actuator to the bracket on the Boom and insert pin Insert keeper...

Страница 6: ...ms Then use the bolt to attach and fasten the L shaped arms to the Mast L shape arms Bolt Bolt KNEEPAD ASSEMBLY FOR STS600E STS600E D BATTERY PLUG INS FOR E series 6 Attach actuator and pendant to the...

Страница 7: ...tery indicator LCD display LCD Lifting actuator UP Lifting actuator DOWN Emergency button EMERGENCY STOP FOR STS600E series LCD INDICATORS IT S STRONGLY RECOMMENDED THAT THE BUTTON IS PRESSED DOWN WHE...

Страница 8: ...y to bear some weight and exibility in knees hips and ankles OPERATING INSTRUCTIONS PREPARATION BEFORE LIFTING Turn on the power by twisting the RED RESET BUTTON clockwise Press the UP or DOWN button...

Страница 9: ...g ConvaQuip Stand Assist Sling Push lift towards patient Open the base of the lift Apply the brakes in both rear casters Position patient s feet on the foot platform and knees against the knee pad Att...

Страница 10: ...patient s knees are locked in a standing position At the standing position patient s knees may move away from the knee pad Release the brakes close the base and pull the lift away from the chair Lowe...

Страница 11: ...essary items i e slings wheelchairs etc are ready for use At least once a month the lift should be thoroughly inspected by a person quali ed to recognize any signs of wear and tear and looseness of bo...

Страница 12: ...sure the hand control is properly inserted all the way into the control box port Check to make sure the actuator plug is properly connected and inserted all the way into the control box TROUBLESHOOTIN...

Страница 13: ...ncluding casters control box PCB pendant charger 2 years Batteries 1 year Weigh Scale 2 years Stand Aid all components 2 years Hydraulic pumps 1 year Reusable slings 6 months ConvaQuip Single Patient...

Отзывы: