CS6103 Interferential Stimulator
Patient Instruction Manual
Control Solutions LLC
Page 28 of 36
Modes of Operation
This section of the document lists the modes of operation and describes what is
generally accomplished by each mode.
For each mode Table 9 lists the treatment time, default frequencies and default sweep
duration.
Mode
Treatment
Time
default
Frequency 1
(Low Beat)
default
Frequency 2
(Highest Beat)
default
Sweep
Time
default
Std
– Standard
20 min.
†
1 Hz
†
10 Hz
6 sec.
36R
30 min.
†
-30% of set
frequency
+60% of set
frequency
6 sec.
36A
60 min.
†
-30% of set
frequency
+60% of set
frequency
6 sec.
Cnt
– Continuous
60 min.
†
100 Hz
†
n/a
n/a
Stm
– Muscle
Stimulation
20 min.
†
48 Hz
†
n/a
n/a
Mem
Table 9
- Mode Details
Table 10 provides a listing of the modes of operation and a description of each.
Mode
Description
Standard (Std)
Provides a beat frequency of 1Hz for 6 seconds followed by a
beat frequency of 10Hz for 6 seconds. This cycle repeats for half
the configured treatment time. Next a beat frequency that
sweeps from 80 to 150Hz within 6 seconds occurs and
continuously repeats for remaining treatment time. The unit
turns off after the treatment period.
36R
Provides a smooth ramp from 30% below the set frequency to
60% above set frequency over sweep time, then back to 30%
below set frequency.
36A
Provides an output frequency set at 30% below the set
frequency for sweep time, then jumps to 60% above set
frequency for sweep time, then repeats.
Continuous (Cnt)
Provides a beat frequency as set on the display for the full
treatment time. After the auto timer timeout out, the unit will turn
off to conserve battery power.
Programmed treatment by clinician
†
These parameters can be changed from the default using the menu.