SPECIFICATIONS

Ranges:

 

0 to 199.9 uW/cm

2

 

0 to 1.9999 mW/cm

2

 

0 to 19.99 mW/cm

2

Resolution:

 

0.1 uW/cm

2

 (0 to 199.9 uW/cm

2

)

 

0.001 mW/cm

2

 (0 to 1.9999 mW/cm

2

)

 

0.01 mW/cm

2

 (0 to 19.99 mW/cm

2

)

Accuracy:

 

±2% full scale plus 2 digits

Wide band 

wavelength: 

320 to 390 nm

Sampling Time:

  Approximately 0.4 seconds

Data Output:

 

PC serial interface

Case:

 

ABS plastic

Fall safe:

 

Low battery indicator

Power:

 

One (1) 9-Volt alkaline battery

Accessories 

Supplied:

 

carrying case, UV sensor probe, 

battery, desiccant Traceable

®

 

Certificate, instructions.

FIGURE 1:

DESCRIPTION

3-1 LCD 

3-2 

Power Button

3-3 

Hold Button: press to hold reading

3-4 

Record Button: press to record results

3-5 

Recall Button: press to show minimum  

 

and maximum readings

3-6 

Zero Button

3-7 

Range Switch

3-8 

UV Sensor

3-9 

Sensor Plug

3-10 

Sensor Input Socket

3-11 

Computer Output Socket

3-12 

 Battery Compartment

UV LIGHT METER OPERATION

1.  Turn the meter on by pressing the POWER BUTTON 

(3-2, fig. 1).

2.  Zero Adjustment: Set the RANGE SWITCH  

(3-7, fig. 1) to 199.99. This is the far left setting. Cover 

the UV SENSOR (3-8, fig. 1) with hand to protect 

it from the environment for a more accurate zero 

reading. Press the ZERO BUTTON (3-6, fig. 1). The 

display should read 0.0. Remove your hand. 

NOTE:

 if the display shows: “CALO” the further 

adjustment is needed. Turn the power off by pressing 

the POWER BUTTON. Press and hold the ZERO 

BUTTON. Press the POWER BUTTON again. The 

bottom line of the display will show “0000”. Adjust the 

VR9 by turning the knob on the right of the unit until 

the top line of the display reads “0”.

3.  Select the desired range by switching the RANGE 

SWITCH (3-7, fig. 1) 

NOTE:

 If the display shows “----” 

then there is too much light for the meter to read in 

this range. Select the next higher range.

4. 

Place the UV SENSOR (3-8, fig. 1) directly under the 

light source you wish to measure. Hold for at least 0.4 

seconds. The meter will measure the light and display 

the value. Limit the angle of light from the source 

being measured to 45 degrees; if possible, limit the 

angle of light to within 30 degrees.

5.  To hold a measurement on the display, press the 

HOLD BUTTON (3-3, fig. 1) while a measurement is 

being taken. The LCD will show 

DH

 in the upper left 

portion of the display to indicate that the value is a 

“held” value. To cancel the data hold feature, simply 

press the HOLD BUTTON a second time.

6.  To record a measurement, press the RECORD  

BUTTON (3-4, fig. 1) while a measurement is being 

taken. The LCD will show 

REC

 in the lower left por-

tion of the display to indicate that the value is being 

recorded. To deactivate the record function, press the 

RECORD BUTTON again.

7. 

Memory Recall: Use the RECALL BUTTON (3-5, fig. 

1) to recall the minimum and maximum readings. 

When the 

REC

 symbol appears or the display, 

press the RECALL BUTTON (3-6, fig. 1) once. The 

maximum recorded value will be displayed. The 

letters “Max” will also appear on indicating that this 

is the maximum or highest reading. A second press 

of the RECALL BUTTON will display the minimum 

reading. The letters “Min” will also appear indicating 

that this is the minimum lowest reading.

Note:

 The Data Record function must be in use to 

utilize the Memory recall features. Once the RECORD 

BUTTON has been pressed a second time to 

deactivate the data record function, the minimum and 

maximum values are no longer stored.

www.

GlobalTestSupply

.com

Find Quality Products Online at:

[email protected]