CARING FOR THE UV LIGHT METER
It is recommended that the meter be recalibrated every
year. The sensor is extremely sensitive to humidity. Store
in a low humidity environment by keeping it in its case
with the Humidity Sponge
™
provided. Proper storage
will extend the length of time between recalibration of
the unit.
COMPUTER SERIAL INTERFACE
This unit features computer output. A COMPUTER
OUTPUT SOCKET (3-11 fig. 1) is located on the top
of the unit. To utilize this feature, connect the unit to a
PC with an accessory Data Acquisition System.
See
Accessories
.
Data is displayed in a 16 digit stream: D15 through D0.
Each digit indicates the following:
D0
end word
D1 to D4 Upper Display Reading, D1=LSD, D4=MSD
D5 to D8 Lower Display Reading, D5=LSD, D8=MSD
D9
Decimal Point (DP) for Upper Display:
0= no DP, 1= 1 DP, 2=2 DP, 3=3 DP
D10
Decimal Point (DP) for Lower Display:
0= no DP, 1= 1 DP, 2=2DP,3=3DP
D11/D12 Annunciator for Upper Display:
00= No Symbol, 01= °C, 02= °F, 03= %,
04= %RH, 05= %PH, 06= %O
2
07= mg/L,
08= m/s, 09= knots, 10= Km/h, 11= Ft/min,
12= mph, 13= µS, 14=mS, 15= Lux,
16= Ft-cd, 17=dB, 18=mV
D13
Annunciator for lower display: 0= No symbol,
1= 00., 2= °F
D14
Reading Polarity for the display:
0= both upper and lower displays are ~
1= upper is “-” and lower is “+”
2= upper is “+” and lower is “–”,
3= both upper and lower displays are “–”.
D15
Start Word
BATTERY LIFE
If the letters “LBT” appear on the left corner of the
display, it indicates the batteries are low and need to
be replaced. To replace the battery, slide the battery
cover located on the back of the unit away from the unit.
Remove the old battery and replace it with a new 9-Volt
alkaline battery. Use an alkaline battery, NOT a regular or
heavy duty battery. Property connect the battery. Replace
the battery cover. Incorrectly installed batteries may
damage electronics.
The unit has an automatic shut off feature to prolong
battery life. If no button on the unit is pressed for ten
minutes, the unit will automatically shut off. To deactivate
this feature, press the RECORD BUTTON while a
measurement is being taken.
ALL OPERATIONAL DIFFICULTIES
If this Ultra Violet Light Meter does not function properly
for any reason, please replace the battery with a new
9-Volt alkaline battery (see Low Batt section, above).
Low battery power can occasionally cause any number o
“apparent” operational difficulties. Replacing the battery
with a new fresh battery will solve most difficulties.
ACCESSORIES
Cat. No. 3150 Humidity Sponge™ Indicator Desiccant
(40 bags/pack)
Contains a blue indicating gel, which turns to pink when
the desiccant pack has achieved its moisture capac
-
ity and needs to be replaced. The desiccant (sodium
calcium aluminosilicate hydrate) is nontoxic and is a
“Generally Recognized as Safe” (GRAS) material. Packs
are certified as Class 100 Clean-Room compatible. Each
pack is 3 x 3 x ¼ inches.
Cat. No. 4138 Easy-Use Accessory Adaptor 115 VAC
Bonus Free Disk This demonstration disk shows the data
acquisition capabilities of this digital product when con-
nected to a computer. See Cat No. 4136 below to order
complete Data Acquisition System.
Cat. No. 4136 Data Acquisition System Accessory
Powerful and easy to use computer data capture/data
logging program works with Traceable
®
Instruments with
computer output. Records interval readings from 1 to
10,000 seconds; displays minimum/maximum readings;
and utilizes an alarm mode that permits the user to be
notified visually, audibly, and by email when an alarm is
triggered. Data is stored to a file that can be printed in
any report or spreadsheet format. Networking server/
client capability allows the captured data to be monitored
on a remote workstation and/or by email. It is designed to
work with Windows
®
98/Me/NT/2000/XP/Vista. Includes
a CD, a 6-foot cable (supplied USB and serial con-
nections) that plugs into the instrument and computer.
Accessory extension cables expand cable length to 300
feet.
Cat No. 4325 Data Logger
Complete DAS–4™ System
captures and stores up to 8000 bytes (over 1000
readings) from all meters. Reading may be taken at
intervals from 1 second to 99 hours. Stored readings
may be downloaded to any PC and viewed. Can be read
“as is” or imported to spreadsheets, databases, and
statistical programs. Supplied with 36-inch serial cable
with D9F computer plug, disk (Windows
®
), and four AA
alkaline batteries. Size is 5 x 3 x 1 inches. Weight is 7
ounces.
Cat. No. 4326 Accessory Adaptor 115 VAC for
Data Logger
www.