the condensing unit is 208-230 volts in which the wiring includes a neutral and a
mechanical ground. This wiring should be connected to the appropriate power
source by a qualified electrician and must conform to all local electrical codes.
SPECIAL VOLTAGE CONNECTIONS
When models are ordered from the factory with special, optional voltages,
connections should be made as required on the electrical wiring diagram provided
on the upper cabinet end panel next to the electrical console box.
START-UP CHECKLIST
After your model has been installed, leveled, cleaned, and electrically connected in
accordance with this manual, please take the time before start-up to observe the
following precautions to assure troublefree operation:
1. Check that all exposed refrigeration lines are free of severe dents or kinks.
2. Check the condenser fan and evaporator fans for freedom to rotate without
any obstructions
3. Make sure that the cabinet is properly leveled (see "leveling" under
installation and location section).
The system should run smoothly and quietly in accordance with generally
accepted commercial standards. If any unusual noises are heard, turn the unit off
immediately and check for any obstructions of the condenser or evaporator fans.
Fan motors, fan blades, or fan housings can be jarred out of position through
rough handling in transit or during installation.
Important Note:
All freezer models have an energy-saving fan delay control to
prevent the interior fans from operating until the evaporator becomes chilled.
Therefore, on initial start-up, the fans will not turn immediately.
This is a normal
mode of operation.
Allow sufficient time for the evaporator coil to reach its
operating temperature (approximately 20 minutes).
Caution:
If unit is unplugged or disconnected for any reason, allow several
minutes (5-6 minutes) before turning the unit back on to allow the system
pressures to equalize. Disregarding this procedure could cause an overload and
prevent the unit from operating.
REMOTE APPLICATIONS
All models are available for purchase as remote models in which case the
condensing unit is purchased separately and installed at the time of installation. All
remote models are equipped with an expansion valve located within the evaporator
coil housing, and both liquid and suction lines stubbed and extending out from the
cabinet case top behind the front grill. Installation of the refrigeration accessories,
- 10 -
Содержание Refrigerators and Freezers
Страница 3: ...TROUBLESHOOTING AND SERVICING GUIDE 22 WIRING DIAGRAMS 26...
Страница 29: ...WIRING DIAGRAMS 26...
Страница 30: ...27...
Страница 31: ...28...
Страница 32: ...29...
Страница 33: ...30...
Страница 34: ...31...
Страница 35: ...32...
Страница 36: ...33...