HD Ultra X Series Instructions
5
8.
Retighten the 3 x T10 locking
screws to hold the Lid Cover
down and in place.
9.
Press down on the Guide Plate
until both ends lock in place to
close.
10.
With the scanner powered up,
and the OTAC pushed down,
with the added +1 click on the
ratchet, make sure the LED for
the raised lid is off.
11.
If the LED does NOT go out
when the OTAC is closed, then
remove the rear panel to access
the sensor, to make adjustments.
Remove the T10 screws and lift
the rear panel off from the back
of the scanner.