The Pulse Oxygen Saturation
(
SpO
2
)
2-digit digital
OLED display
Pulse Rate
(
PR
)
3-digit digital
OLED display
Pulse Intensity (bar-graph)
bar-graph
OLED display
Waveform
Waveform
OLED display
SpO
SpO
SpO
SpO
2
2
2
2
Parameter
Parameter
Parameter
Parameter Specification
Specification
Specification
Specification
Measuring range
0%
~
100%, (the resolution is 1%).
Accuracy
70%
~
100%
:
±2% ,Below 70% unspecified.
Pulse
Pulse
Pulse
Pulse Parameter
Parameter
Parameter
Parameter Specification
Specification
Specification
Specification
Measuring range
30bpm
~
250bpm, (the resolution is 1bpm)
Accuracy
±2 bpm or ±2% (select larger)
Safety
Safety
Safety
Safety Type
Type
Type
Type
Interior Battery
,
BF Type
Pulse
Pulse
Pulse
Pulse Intensity
Intensity
Intensity
Intensity
Range
Continuous bar-graph
display, the
higher display
indicates the stronger pulse.