discharge
(ESD)
±8KVair
±8KVair
concrete or ceramic tile. If
IEC 61000-4-2
floor are covered with
synthetic material,
the
relative humidity should be
at least 30%.
Power frequency
Mains power quality should
(50Hz) magnetic
be that of a typical
3Nm
3Nm
field
commercial or hospital
IEC 61000-4-8
environment
NOTE: UT is the a.c. mains voltage prior to application of the test level.
Guidance and manufacture's declaration-electromagnetic immunity
for EQUIPMENT and SYSTEMS that are not LIFE-SUPPORTING
Guidance and manufacture's declaration-electromagnetic immunity
The ECG80A Electrocardiograph (ECG) is intended for use in the electromagnetic environment
specified below. The customer or the user of the ECG80A Electrocardiograph (ECG) should
assure that it is used in such an environment
Immunity
IEC6060
I
Complianc
I
test e level
I
Electromagnetic environment -guidance
test
level
Conducted
3Vrms
3Vrms
Portable and mobile RF communication equipment
RF
l50KHz
should be used no closer to any part of the ECG80A
ICE
to
Electrocardiograph (ECG), including cables, than the
61000-4-6
80MHz
recommended separation distance calculated from the
equation applicable to the frequency of the transmitter.
recommended separation distance
3V/m
I
3V/m
d
= [3.5}/p
I
Vl
Radiated
80MHz
RF
to
d
=[ �� ]"1>
IEC6I000-4
2.5GHz
80MHz to 800MHz
-3
28