Appendix 2
Guidance and manufacture's declaration-electromagnetic emission
for all EQUIPMENT and SYSTEMS
Guidance and manufacture's declaration -electromagnetic emission
The ECG80A Electrocardiograph (ECG) is intended for use in the electromagnetic environment
specified below. The customer of the user of the ECG80A Electrocardiograph (ECG) should
assure that it is used in such an environment
Emission test
compliance
Electromagnetic environment-guidance
The ECG80A Electrocardiograph (ECG) uses RF
energy only for its internal function. Therefore, its
RF emissions
Group l
RF emissions are very low and are not likely to
CISPR 11
cause any interference in nearby electronic
equipment.
RF emissions
ClassB
The ECG80A Electrocardiograph (ECG) is
CISPR l l
suitable foe use in all establishments, including
Harmonic emissions
Not applicable
domestic establishments and those directly
IEC 61000-3-2
connected to the public low-voltage power supply
Voltage fluctuations/
network that supplies buildings used for domestic
flicker emission
Not applicable
purposes.
IEC 61000-3-3
Guidance and manufacture's declaration-electromagnetic immunity
for all EQUIPMENT and SYSTEMS
Guidance and manufacture's declaration-electromagnetic immunity
The ECG80A Electrocardiograph (ECG) is intended for use in the electromagnetic environment
specified below. The customer or the user of ECG80A Electrocardiograph (ECG) should assure
that it is used in such an environment
Electromagnetic
Immunity test
IEC6060 l test level Compliance level
environment-guidance
Electrostatic
±6KV contact
±6KV contact
Floors should be wood,