1
C861 • C862 • C863 • C864 • C868
C861 • C862 • C863 • C864 • C868
2
Inputs
The pH/mV/Ion electrodes should be connected to the coaxial pH/mV connectors
CH.1, etc. If separate electrodes are used, connect the reference electrodes to the
REF. terminals. Connect the conductivity cell to the coaxial EC input. Automatic
temperature compensation and temperature measurements are possible by plug-
ging a Pt1000 thermocompensator into the °C terminals. Without compensator, the
manual temperature compensation is automatically switched on. C862 & C863: the
O
2
-electrode should only be connected to the coaxial connector CH.2. Its banana
plugs should be inserted in the °C terminals.
Password
1.
Select [MENU] by pressing
MODE
.
2.
Select [PASSWORD] and press
CAL
.
3.
A private code can be programmed to avoid undesired access to the instru-
ment. Choose [YES] and press
CAL
to enter your secret sequence of 5 keys.
Recorder output
A recorder can be connected to the red (+) and black (-) terminals. Use only labora-
tory recorders with a high input impedance!
Computer interface
A USB port is provided for interfacing the instrument with a computer. Download
the necessary driver from www.consort.be (>downloads >software).
System
1.
Select [MENU] by pressing
MODE
.
2.
Select [SYSTEM] and press
CAL
.
3.
Follow the instructions on the screen to adjust language and contrast.
Date and Time
1.
Select [MENU] by pressing
MODE
.
2.
Select [DATE/TIME] and press
CAL
.
3.
Adjust the date and press
CAL
to adjust the date.
4.
Adjust the time and press
CAL
to adjust the time.
Recorder
1.
Select [MENU] by pressing
MODE
.
2.
Select [RECORDER] and press
CAL
.
3.
Select the desired channel and press
CAL
.
4.
Select the desired range and press
CAL
(only for conductivity).
5.
Select the desired minimum level corresponding to a 0 V recorder output and
press
CAL
.
6.
Select the desired maximum level corresponding to a 2 V recorder output and
press
CAL
.
Digital output
A standard RS232 output terminal (DP9) is provided for interfacing the instru-
ment with a printer or computer. Data is sent in the ASCII code at a BAUD rate of
300...19200 bps (8 bit, no parity, 1 stopbit).
Serial port pinout specifications:
pin 1 : connected to pin 4 and pin 6
pin 2 : TxD, transmit data
pin 3 : RxD, receive data
pin 4 : connected to pin 1 and pin 6
pin 5 : Gnd, signal ground
pin 6 : connected to pin 1 and pin 4
pin 7 : connected to pin 8
pin 8 : connected to pin 7
pin 9 : not connected