4-6
4.3.1 CAL Mode General
This unit may be completely calibrated without
selecting or adjusting components. Parameters
necessary to compensate for unit-to-unit manufac-
turing tolerances and to normalize performance
to specification are stored digitally in a non-
volatile memory device on the A3 Controller
PWB. By use of external calibration standards and
the unit's menu-driven CAL Mode, calibration
may be completed quickly using only the normal
controls and displays.
To enter CAL Mode:
1. Turn unit off. Open the top cover and set
A3S1, accessible through a hole in the shield, to
CAL Mode. Make sure shield is in place and close
top cover.
2. Connect unit to an AC source of the nameplate
rating. Connect a dual treadle foot switch.
3. Power up while pressing the PAD button. If a
HLP 9 is displayed, the unit did not detect the
PAD button closure. Power down, or cycle A3S1
to RESET and back to CAL, and try again.
4. Normal entry to CAL Mode is signalled by a
continuous Lamp Test, with all display elements
steadily lit.
5. Release and press PAD again. The "CC" menu
selection will appear in the Coag window, and
300, 400 or 500 will appear in the Cut (Option)
window. The Menu selections and options may be
stepped through using the Cut and Coag Power
Control knobs. See Fig 4.4 for available CAL
Mode Menu and option selections.
The following sections contain only step-by-step
calibration procedures. If difficulty is encoun-
tered, refer to the detailed description of CAL
Mode theory and software in Section 3.5.2.
Upon completion of calibration, ensure that the
shield is securely installed, switch A3S1 is in
RUN Mode, and the Top Cover is secured with
both rear screws. Power up in RUN Mode to ver-
ify that a HLP 3 (NOVRAM CRC) fault is not
declared before returning the unit to service.
MODE
MODE USE
OPTION
OPTION USE
(COAG PWR)
(CUT PWR)
CC
MONOPOLAR
300, 400, 500
SELECT LOAD
POWER CAL
RESISTANCE FOR
MONOPOLAR CAL
bP
BIPOLAR PWR CAL
50, 100, 125
SELECT LOAD
RESISTANCE FOR
BIPOLAR CAL
Pd
ARM RESISTANCE
BLANK
CALIBRATION
dI
DIAGNOSTICS
1
WATCHDOG
TIMER CHECK
2
PSEUDO RUN MODE
Table 4.4 CAL Mode Menu