3.2
Troubleshooting
Table 4: Troubleshooting
Symptom
Possible Cause
Corrective Action
Handpiece does not
operate.
♦
Handpiece cord not connected
securely.
♦
Handpiece cord and/or button
faulty.
♦
Handpiece faulty.
♦
If using a footswitch, cord not
connected securely.
♦
Footswitch cord or footswitch
faulty.
♦
Handpiece safety is in the safe,
or off position.
♦
Twist collet not in the fully
locked position.
♦
Securely connect handpiece cord
to handpiece and controller.
♦
Replace handpiece cord.
♦
Return for service.
♦
Securely connect footswitch cord
to controller receptacle.
♦
Return for service.
♦
Move safety to the appropriate
operating position.
♦
Twist collet to the fully locked
position.
Handpiece overheats
♦
Worn bearings in the bur
guard.
♦
Moisture in the handpiece.
♦
Reference “3.3 Bur Guard Test-
ing Procedure” on page 26. If
resistance is noticed, return bur
guard for service. If no resistance
is noticed, return for service.
♦
Resterilize the handpiece accord-
ing to sterilization parameters and
drying times, reference “3.1.4
Sterilization Information” on
page 23.
“Magnetic Field”
is displayed on the
controller.
♦
The handpiece is in close
proximity to a magnetic drape.
♦
Removable activation lever
partially connected.
♦
Faulty handpiece.
♦
Remove the handpiece from the
magnetic drape area. Wait five
seconds before restarting.
♦
Completely connect the activation
lever (See “2.3.4 Removing the
Detachable Handpiece Activa-
tion Lever” on page 16).
♦
Return handpiece to ConMed
Linvatec for service.