EVENT LOG
The Event Log provides a record of any events that may occur while the IntelliSwitch is running. This
can be an alarm condition or a key press. Essentially, any “event” that may occur either as a result of a
response to a change in conditions or operator intervention. A list of “events” that could be displayed in
the event log is shown below.
Inlet Value Status
– “description”
Error
– “description”
Left Side O/P
– “value” (O/P = over pressure)
Right Side O/P
– “value”
Reserve Alarm
Left Side U/P
– “value” (U/P = under pressure)
Right Side U/P
– “value”
Reserve In-Use
Keypad Lock
“description”
Cylinder Selection Changed
– “description”
Delivery Regulator O/P
– “value”
Delivery Regulator U/P
– “value”
Delivery Regulator Status
– “value”
Reserve Status
– “value”
Alarm Clear U/P Left Side
– “value”
Alarm Clear O/P Left Side
– “value”
Alarm Clear U/P Right Side
– “value”
Alarm Clear O/P Right Side
– “value”
Alarm Clear Reserve Gas
The word “value” is replaced by a pressure in psi. The word “description” is replaced with and
explanation of what happened.
The Event Log can store up to 50 events. Once the quantity of events reaches 50 it will continue to list
the events but push events off the list starting with the first one. The list will be maintained until it is
cleared. To clear the Event Log, click on the “CLEAR” button.
The Event Log can also be printed if a permanent record is needed. The PRINT function can use
whatever printer the user’s computer is connected to. To print the event log, click on the PRINT button.
A normal windows print screen will appear.