26
27
4. * Caution: This machine should not be used adjacent to or stacked with other
equipment, and that if adjacent or stacked use is necessary, this machine should be
observed to verify normal operation in the confi guration in which it will be used.
Guidance and manufacturer’s declaration – electromagnetic immunity
The Lumilisse Hair Remover (IPL960F) is intended for use in the electromagnetic
environment specifi ed below. The customer of the user of the Lumilisse Hair
Remover (IPL960F) should assure that it is used in such an environment.
Emission test
Compliance
Electromagnetic environment – guidance
RF emissions
CISPR 11
Group 1
The Lumilisse Hair Remover (IPL960F) use
RF energy only for its internal function.
Therefore, its RF emissions are very low and
are not likely to cause any interference in
nearby electronic equipment.
RF emission
CISPR 11
Class B
The Lumilisse Hair Remover (IPL960F)
is suitable for use in all establishments,
including domestic establishments and those
directly connected to the public low-voltage
power supply network that supplies buildings
used for domestic purposes.
Harmonic
emissions
IEC 61000-3-2
Class A
Voltage
fl uctuations/
fl icker
emissions
IEC 61000-3-3
Complies
Guidance and manufacturer’s declaration – electromagnetic immunity
The Lumilisse Hair Remover (IPL960F) is intended for use in the electromagnetic
environment specifi ed below. The customer or the user of Lumilisse Hair Remover
(IPL960F) should assure that it is used in such an environment.
Immunity
test
IEC 60601
test level
Compliance
level
Electromagnetic environment –
guidance
Electrostatic
discharge
(ESD)
IEC 61000-4-2
±8 kV contact
±2 kV, ±4 kV,
±8 kV, ±15
kV air
±2 kV for
power supply
lines
± 1 kV line(s)
to line(s)
<5% UT
(>95% dip in
UT)
for 0.5 cycle
±8 kV contact
±2 kV, ±4 kV,
±8 kV, ±15
kV air
±2 kV for
power supply
lines
± 1 kV line(s)
to line(s)
<5% UT
(>95% dip in
UT)
for 0.5 cycle
Floors should be wood, concrete
or ceramic tile. If fl oors are covered
with synthetic material, the
relative humidity should be at
least 30%.
Electrical fast
transient/
burst
IEC 61000-4-4
40% UT
(60% dip in
UT)
for 5 cycles
40% UT
(60% dip in
UT)
for 5 cycles
Mains power quality should be
that of a typical commercial,
hospital, or home environment.
Surge
IEC 61000-4-5
70% UT
(30% dip in
UT)
for 25 cycles
70% UT
(30% dip in
UT)
for 25 cycles
Mains power quality should be
that of a typical commercial,
hospital, or home environment.
Voltage
dips, short
interruptions
and voltage
variations on
power supply
input lines
IEC 61000-
4-11
<5% UT
(>95% dip in
UT)
for 5 sec
<5% UT
(>95% dip in
UT)
for 5 sec
Mains power quality should be
that of a typical commercial,
hospital, or home environment.
If the user of the Lumilisse Hair
Remover (IPL960F) requires
continued operation during
power mains interruptions,
it is recommended that
the Lumilisse Hair Remover
(IPL960F) be powered from an
uninterruptible power supply
or a battery.
Power
frequency
(50Hz/60Hz)
magnetic fi eld
IEC 61000-4-8
30 A/m
30 A/m
Power frequency magnetic fi elds
should be at levels characteristic
of a typical location in a typical
commercial, hospital, or home
environment.
NOTE:
UT is the AC’s main voltage prior to application of the test level.