7.0 GAS CYLINDER LOW VOLUME MEASUREMENT
7.1 Description
The GTFM can be used to provide the user with an indication of when the volume in
a cylinder of gas is reaching a low level. To use this function the user must set the
starting volume in the cylinder and a minimum value for testing. When the minimum
value is reached the GTFM will flash a
"G"
on the display. The GTFM will display the
flashing
"G"
until the original volume is zero. To disable this function set the
"V"
parameter to zero.
7.2 Setup
To set the starting volume select the
"V"
parameter by consecutively pressing the
"MODE"
switch to the up (
"SET")
position until the
"V"
parameter is displayed. While
viewing the
"V"
parameter press the
"MODE"
switch to the down (
"PGM")
position to
increment to the desired value. To set the minimum limit select the
"M"
parameter by
pressing the
"MODE"
switch consecutively to the up (
"SET")
position until the
"M"
parameter is displayed. While viewing the
"M"
parameter press the
"MODE"
switch
to the down (
"PGM")
position to increment to the desired value.