Statox 503 Control Module
Issue 10 / 2018 Page
9
of
46
If a sensor head has been connected and the appropriate program has been selected, the measuring mode
will start.
Sensor heads for combustible gases, type HRC, ARE, LC IR, MCIR, CO
2
and PID must be
calibrated after being connected.
Exception: A line calibration has been done by Compur.
The message „CALIB. REQUIRED“ alternating with the measured value, tells you that the
combination sensor head / control module has not been calibrated together. The control
module operates with default sensor parameters. By pushing ENTER, the message „CALIB.
REQUIRED“ can be blanketed. By pushing ENTER in the measuring mode, this message can
be recalled: if there was no initial calibration, “NO VALID CAL. DATA“ is shown for 5s.
!
Power ON
Actual firmware version is displayed 20 s
Actual operation program is displayed 5 s.
During this time an access to the main menu is possible.
Figure 5: Start-up sequence
Measuring Mode