background image

 

 

16 

 

Initial Checkout Procedure 

 

The  following  procedure  will  verify  that  the  MegaPulse  1.2x50-16PF  PV  tester  is  working 
correctly.    We  recommend  that  this  procedure  be  conducted  periodically  to  ensure  proper 
operation  of  the  tester.    A  measuring  instrument  to  monitor  the  output  waveform  is  needed  to 
conduct  this  procedure.    Always  ensure  that  the  measuring  instrument  (usually  an  oscilloscope 
with a high-voltage probe) is  rated for the voltage involved, and that the frequency response of 
the  instrument  and  probe  are  capable  of  measuring  the  output  waveform  of  the  MegaPulse 
1.2x50-16PF  PV  tester.    A  measuring  instrument  or  probe  with  a  low  frequency  response  will 
result in erroneous readings that could be mis-read.  

  

CAUTION 

High voltage generated by the MegaPulse tester is exposed during this test.  A 

risk of shock exists.  Exercise care when using the MegaPulse tester. 

 

1. 

Connect the tester to a proper source of supply using the included 18 AWG power supply 

cord.   

2. 

Connect the red and black test leads to the EUT. 

3. 

Connect  the  ends  of  the  test  leads  to  an  appropriate  measuring  instrument  (typically  an 

oscilloscope with a high-voltage probe).   

 
 

NOTE:    The  output  of  the  MegaPulse  is  a  high-energy  fast-rise  (>10kV/uS)  waveform.  

Depending on the measurement setup, it is possible to see very large common-mode noise 
which  can  be  mis-interpreted  as  high-frequency  noise  on  the  front  edge  of  the  impulse 
waveform.  Figure 5 shows a test setup to measure common-mode noise.  The high-voltage 
oscilloscope probe, and the probe ground are both connected to one side of the MegaPulse 
output.    In  theory,  the  result  should  be  that  0  volts  is  measured  on  the  oscilloscope.    In 
practice,  because  of  the  high-impedance  of  the  high-voltage  oscilloscope  probe,  the 
oscilloscope  will  measure  a  substantial  high-frequency  transient  as  shown  in  Figure  6.  
Before  making  measurements  of  the  actual  impulse  output  waveform,  be  sure  to  perform 
some measurements as shown in figure 5 and take steps to minimize the measured noise as 
much as possible.  Some possible remedies to reduce common-mode noise are as follows: 
 

a)

 

Use a suitable oscilloscope and high-voltage probe. 

b)

 

Use a battery-powered oscilloscope. 

c)

 

Isolate the oscilloscope from ground. 

d)

 

Adjust the scope to limit the measurement bandwidth. 

e)

 

Perform  a  two-channel  differential-mode  measurement,  where  the  common-mode  noise 
is subtracted from the impulse waveform. 

 
 

If this common-mode noise is not eliminated in the test setup, the common-mode noise will 

also be measured when the high-voltage impulse output is measured.  The result will look 
like  the  waveform  shown  in  Figure  7,  where  the  high-voltage  common-mode  noise  is 
essentially added to the differential (impulse) voltage.   

Содержание MegaPulse

Страница 1: ...MegaPulse IMPULSE TESTER 1 2x50 16PF PV Instruction Manual...

Страница 2: ...gn and ergonomic front panel and represents the latest in high voltage impulse testing To fully appreciate all the features of your new instrument we suggest that you take a few moments to review this...

Страница 3: ...3...

Страница 4: ...Testing Area 6 Safety Techniques 7 Using the MegaPulse Impulse Tester 7 Introduction 9 Operation 11 Setting up your Tester 11 AC Line Voltage Requirements 11 Fuse Replacement 12 Front and Rear Panel...

Страница 5: ...5...

Страница 6: ...re that high voltage is adept and creative at completing a path to ground Instructions should include a warning against any metal jewelry Operators should not allow others in the testing area especial...

Страница 7: ...er test from the MegaPulse until the front panel meter reads a safe level 40 volts or less is generally considered a safe level Pressing the TRIGGER button before disconnecting main power or turning t...

Страница 8: ...5 Ohms and the MegaPulse is capable of delivering thousands of Amps of current into a short circuit While the MegaPulse can withstand a short circuit at the output terminals it should NOT be used in t...

Страница 9: ...Discharge progress is shown on the front panel meter Your tester is warranted for a period of one year upon shipment of the instrument to the original purchaser Specifications Output Peak 800V 16kV To...

Страница 10: ...follow test sequences varying test voltage and time between tests and keep a record with test information Compatible with Windows 32bit platform 220 230 or 240 If a different line voltage is desired...

Страница 11: ...e use the original shipping container If the original shipping container is not available be sure that adequate protection is provided to prevent damage during shipment We recommend that the instrumen...

Страница 12: ...h to the O or off position 2 Unplug the instrument from the source of supply 3 Remove the power inlet cord from the instrument 4 Using a small screwdriver pry open the fuse holder door 5 Replace the f...

Страница 13: ...13 Figure 3 Controls Indicators Connectors MegaPulse 1 2x50 16PF PV Front Panel...

Страница 14: ...been pressed 5 CHARGE indicator This Yellow indicator is lit to show that pressing the CHARGE switch is the next logical step in a test sequence Pressing the Charge switch causes the CHARGE indicator...

Страница 15: ...ied cordset to connect the MegaPulse tester to appropriate source of supply Fuse holder provides access for Fuse replacement and the Power Switch is used to turn the tester ON and OFF 2 Fuse replaceme...

Страница 16: ...setup it is possible to see very large common mode noise which can be mis interpreted as high frequency noise on the front edge of the impulse waveform Figure 5 shows a test setup to measure common m...

Страница 17: ...17 Figure 5 Common mode noise measurement setup Figure 6 Common mode noise Figure 7 Common mode noise superimposed on the output waveform...

Страница 18: ...nds on the front meter Turn Clockwise to increase the setting Voltage Setting Point 7 Push the yellow CHARGE button Verify the red TRIGGER indicator is now lit 8 After the front panel meter indicates...

Страница 19: ...voltage is present on the main storage capacitor even when the tester is first turned ON This is due to inherent charging of the internal capacitors Pressing the TRIGGER switch will discharge the cap...

Страница 20: ...n Information The Megapulse 1 2x50 16PF PV has been fully calibrated at the factory in accordance to our published specifications It is recommended that you have this instrument re calibrated and safe...

Отзывы: