AC Line Voltage Requirements 

 

AC  line  voltage  requirements  for  your  Tester  are  noted  on  the  rear  panel  of  the  instrument.    Do  not 
connect the instrument to a different voltage source.  The cord packaged with your MegaPulse Tester is 
for use in the United States.  If another power cord must be used, the cord must be rated for the maximum 
current noted on the rear panel.  It must also meet the requirements of IEC 227 or IEC 245, and mains 
cords  that  are  certified  or  approved  by  any  recognized  national  test  house  are  regarded  as  meeting this 
requirement.   

 

Fuse Replacement 

 

There is a user-replaceable fuse (F1) located on the rear panel of the instrument.  It is located behind a 
door in the Power Inlet-Power Switch-Fuse Holder device.  The fuse rating is noted on the rear panel.  Do 
not attempt to replace it with a fuse of any other rating. 
 
Use the following procedure to replace the fuse F1: 

1.  Turn the power switch to the OFF position. 
2.  Unplug the instrument from the source of supply. 
3.  Remove the power inlet cord from the instrument. 
4.  Using a small screwdriver, pry open the fuse holder door. 
5.  Replace the fuse with a new one of the correct rating. 
6.  Replace the fuse holder door and power inlet cord.

   

 

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

Содержание MegaPulse Capacitor A PF Series

Страница 1: ...MegaPulse IMPULSE TESTER Capacitor series A PF Instruction Manual...

Страница 2: ...ign and ergonomic front panel and represents the latest in high voltage impulse testing To fully appreciate all the features of your new instrument we suggest that you take a few moments to review thi...

Страница 3: ......

Страница 4: ...Megapulse Capacitor series A PF Specifications 7 Front and Rear Panel Features 8 Figure 1 Controls Indicators Connectors MegaPulse Capacitor series A PF Front Panel 8 Table 1 Controls Indicators Conn...

Страница 5: ...1...

Страница 6: ...re that high voltage is adept and creative at completing a path to ground Instructions should include a warning against any metal jewelry Operators should not allow others in the testing area especial...

Страница 7: ...d be set up to the left and in front of the equipment under test The equipment under test should be connected to the MegaPulse tester and then left alone by the operator After the operator is clear of...

Страница 8: ...s rated for the voltage involved and that the frequency response of the instrument and probe are capable of measuring the output waveform of the MegaPulse Impulse Tester A measuring instrument or prob...

Страница 9: ...F 1 User Manual 1 High Voltage Test Lead Red 00 HVL7R 3 feet 1 High Voltage Test Lead Black 00 HVL7B 3 feet 1 18 AWG AC Power Cord 70 101 1 TestMinder Optional User Manual 1 Software CD Megapulse Test...

Страница 10: ...l test house are regarded as meeting this requirement Fuse Replacement There is a user replaceable fuse F1 located on the rear panel of the instrument It is located behind a door in the Power Inlet Po...

Страница 11: ...point adjusted by frontal VOLTAGE knob or by PC optional Polarity Control Positive only Voltage Display 4 Digit LED Display Voltage Meter resolution 5V Duty Cycle 1 pulse every 10 seconds Line Voltage...

Страница 12: ...se of its controls indicators and connectors The front panel features of the MegaPulse are shown in Figure 1 and described in Table 1 The rear panel features of the MegaPulse are shown in Figure 2 and...

Страница 13: ...ross the output leads 7 TRIGGER indicator This Red indicator is lit to show that the tester can be trigger TRIGGER indicator is lit for 2 minutes after the CHARGE button is pressed TRIGGER indicator w...

Страница 14: ...d to the chassis of the MegaPulse and is referenced to earth ground as long as the MegaPulse is properly grounded Even though this jack is referenced to ground it should be treated as hazardous whenev...

Страница 15: ...d OFF 2 Fuse replacement warning Rating of power supply Specifies replacement fuse and required supply voltage 3 RS 232 Interface Optional Allow the communication between the tester and computer inter...

Страница 16: ...ect the ends of the test leads to a capacitor load of the range selected on the front panel knob plug your high voltage oscilloscope probe to the output leads as well Note that the RETURN lead is refe...

Страница 17: ...ce West USA is available Phone 800 748 6224 Hours 8 30 AM 4 30 PM Pacific Time Also available on our web site at www compwest com Contact Compliance West USA 650 Gateway Center Way Suite D San Diego C...

Страница 18: ...ll be corrected at no charge Mail the instrument post paid to the manufacturer Dated proof of purchase is required for all in warranty repairs The manufacturer is also available for calibration and or...

Страница 19: ...ation procedure must be performed by qualified personnel for more information contact Compliance West USA Voltage Stop Disable Keyboard Enable by Keyboard If the MegaPulse Capacitor series A PF tester...