Safety Techniques 

 

The high voltage circuit of the MegaPulse can be shut off at any time by turning OFF the rear power switch.  
Note that there are residual voltages inside the equipment that may take up to 85 minutes to bleed off to safe 
levels. Always press TRIGGER to discharge the tester before turning OFF. 
 
The MegaPulse tester is provided with a digital 

VOLTAGE ADJUST

 knob on the front panel.  This voltage 

setting should always be confirmed by pressing the VOLTAGE ADJUST knob before starting any test.   
 
The MegaPulse tester is provided with a 

CHARGE

 switch that is in the unarmed "Standby" setting when the 

tester is first turned ON.  When the yellow 

CHARGE

 button is lit, the tester will not provide high voltage 

until the 

CHARGE

 Button and the 

TRIGGER

 Button have been pressed in order. To prevent inadvertent 

operation, the operator should be instructed not to press the 

CHARGE

 Button until the test is ready. 

 
The MegaPulse tester has been designed for one-touch operation with the right hand.  If possible, it should 
be set up to the left and in front of the equipment under test. The equipment under test should be connected 
to the MegaPulse tester and then left alone by the operator. After the operator is clear of the tester and the 
equipment  under  test,  the  operator  should  turn  the  rear-panel  power  switch  to  ON,  press  the 

CHARGE

 

Button,  adjust  the  voltage  to  the  desired  level  (as  displayed  on  the  front  panel  meter),  then  press  the 

TRIGGER

 Button, with his right hand. This will allow the greatest separation between the operator and the 

test being conducted. 
 
The MegaPulse tester is designed to bleed the high voltage away after the test has concluded. To ensure that 
any voltage present in the equipment being tested has been completely bled away, the operator should not 
unplug the equipment under test from the MegaPulse until the front panel meter reads a safe level (40 volts 
or less is generally considered a safe level). Pressing the TRIGGER button before disconnecting main power 
(or turning the equipment off) will ensure that the internal capacitors are discharged as much as possible. 
 

Safety Markings 

 

Содержание MegaPulse 968-B Metallic A

Страница 1: ...Instruction Manual...

Страница 2: ...i...

Страница 3: ...gn and ergonomic front panel and represents the latest in high voltage impulse testing To fully appreciate all the features of your new instrument we suggest that you take a few moments to review this...

Страница 4: ...iii...

Страница 5: ...ng the Instrument 1 AC Line Voltage Requirements 2 Fuse Replacement 2 Section 3 3 Specifications and Controls 3 Front and Rear Panel Features 4 Section 4 7 Operating Instructions 7 External Interlock...

Страница 6: ...v...

Страница 7: ...nclude a warning against any metal jewelry Operators should not allow others in the testing area especially when tests are being conducted Organization is to be stressed The operator should keep the a...

Страница 8: ...r one touch operation with the right hand If possible it should be set up to the left and in front of the equipment under test The equipment under test should be connected to the MegaPulse tester and...

Страница 9: ...pulse relative to the RETURN jack When the polarity switch is in the Reverse position REV indicator is lit the high voltage will appear on both OUTPUTS as a negative pulse relative to the RETURN jack...

Страница 10: ...748 6224 Product Package for 968 B Metallic A Tester Description Part Number 968 B Metallic A MegaPulse Tester User Manual High Voltage Test Lead Red 00 HVL7R Qty 1 High Voltage Test Lead Black 00 HV...

Страница 11: ...l test house are regarded as meeting this requirement Fuse Replacement There is a user replaceable fuse F1 located on the rear panel of the instrument It is located behind a door in the Power Inlet Po...

Страница 12: ...tage Pulse Front Time 6 s to 10 s 1 67 times between 30 and 90 per TIA 968 B Fig 2 Voltage Pulse Decay Time 560 s to 860 s per TIA 968 B Fig 2 Current Peak 100A minimum Current Pulse Front Time 5 s to...

Страница 13: ...amiliar with the use of its controls indicators and connectors The front panel features of the MegaPulse are shown in Figure 2 and described in Table 2 The rear panel features of the MegaPulse are sho...

Страница 14: ...er is turn ON an after pressing TRIGGER button CHARGE indicator will go out after pressing CHARGE button CHARGE and TRIGGER Indicators will be blinking if the Interlock Switch is open Only testers wit...

Страница 15: ...se replacement warning Rating of power supply Specifies replacement fuse and required supply voltage 3 RS 232 Interface Optional Allow the communication between the tester and computer interface a RS...

Страница 16: ...wn on the computer condition that will also disable polarity changes charge and trigger conditions When the Interlock is closed it enables all normal operations of the MegaPulse features Front Keyboar...

Страница 17: ...Tester is capable to maintain a specific voltage charge selected by the Voltage Set Point CAUTION High voltage generated by the MegaPulse tester is exposed during this test A risk of shock exists Exer...

Страница 18: ...the jacks on the front panel 6 Connect the ends of the test leads to the measuring instrument See Error Reference source not found for an example 7 Set the next capturing setting on the digital oscill...

Страница 19: ...neglect Malfunctions which occur within the limits of the warranty will be corrected at no charge Mail the instrument post paid to the manufacturer Dated proof of purchase is required for all in warr...

Страница 20: ...Pulse but it will not be the same amount it will depend on the waveform calibration and tolerances therefore it is important to measure the peak amplitude of the output waveform and adjust the output...

Страница 21: ...m Compliance West USA is available Phone 800 748 6224 Hours 8 30 AM 4 30 PM Pacific Time Also available on our web site at www compwest com Contact Compliance West USA 650 Gateway Center Way Suite D S...

Отзывы: