14 

Paragraph 8.4.1.3.6.1: Portable 
Equipment Lead to Ground – 
Nonisolated Input Test 

 

1. 

Connect the LCT-601 to the EUT using 
the female attachment plug cap pigtail. 

2. 

Connect the LCT-601 to mains power 
using the male attachment plug cap 
pigtail. 

3. 

Set the MEAS POINT Switch to the 
CHASSIS MEAS POINT (left) position. 

4. 

Regarding the EARTH switch (second 
switch from the top); the test is to be 
conducted with the EARTH switch in 
both positions. 

5. 

Set the NEUTRAL switch (third switch 
from the top) to the NEUTRAL 
CONNECTED position. 

6. 

Set the POLARITY switch to the 
NORMAL position. 

7. 

Set the power switch of the EUT to the 
ON position.  

8. 

Connect a calibrated voltage meter 
across the RED and GREEN LEAKAGE 
banana plugs on the LCT -601.  Set it to 
read ~100uV. 

9. 

Connect all patient-connected leads 
together, connect them to the RED 
LEAKAGE banana plug of the LCT-601 
and conduct measurements for both test 
conditions.  For this test, switch the 
EUT to the output condition that results 
in the highest leakage output. 

10.  Read the current meter for  both 

conditions noted above.  A passing 
result is <100uA (<100uV on the meter 
connected to the LCT-601) for any 
condition. 

11.  See Figure  11 and Table  11  for a Test 

Setup synopsis.  This section is  based 
on NFPA 99 Fig. 8.4.1.3.6.1, see below. 

 

 

Figure 10 Nonisolated Lead Leakage Test 

Table 10 Nonisolated Lead Leakage Test 

Jack/Switch 

Position 

FEMALE Cordset 

To EUT 

MALE Cordset 

To Mains Power 

RED LEAKAGE Jack 

To voltage meter 
and all  nonisolated 
patient connections 
on the EUT 
connected together 

GREEN LEAKAGE Jack 

To voltage meter 

RED RESISTANCE jack 

N/C 

MEAS. POINT Switch 

CHASSIS 

EARTH Switch 

TEST IN BOTH 
POSITIONS 

NEUTRAL Switch 

CONNECTED 

POLARITY Switch 

NORMAL 

EUT Power Switch 

ON 

PASS Result: 

<100uA  

 
 

 

FIGURE 8.4.1.3.6.1 Test Circuit for Measuring 
Leakage Current Between Patient Leads and 
Ground — Nonisolated.

 

Содержание LCT-601

Страница 1: ...LCT 601 NFPA 99 Leakage Current Tester Instruction Manual COMPLIANCE WESTUSA...

Страница 2: ...ckly allow testing for Leakage Current in accordance with the requirements of NFPA 99 Sections 8 4 1 3 for Installed Equipment and NFPA 99 Section 10 2 13 1 for Manufacturer s Tests The operating inst...

Страница 3: ...ii Rev 1 1 May 20 2005...

Страница 4: ...ads Isolated Input 8 LCT 601 for End Users NFPA 99 Section 8 4 1 3 11 Paragraph 8 4 1 3 2 Grounding Circuit Continuity Test 11 Paragraph 8 4 1 3 4 Leakage Current Fixed Equipment 12 12 Paragraph 8 4 1...

Страница 5: ...16 Figure 13 Between Leads Isolated Input 17 Figure 14 Leakage Test for Facility Owned Portable Equipment 18 Table of Tables Table 1 Grounding Circuit Continuity Test 3 Table 2 Leakage Current from Ap...

Страница 6: ...acturers to verify compliance with the requirements of NFPA 99 Section 10 2 13 Section 3 describes use of the LCT 601 for those requirements for Installed Equipment both cord connected and permanently...

Страница 7: ...the left position Switch 4 in right Reverse Polarity position Measuring from plug to receptacle on the cords a Neutral to Neutral open b Hot to Hot open c Hot to Neutral 0 0 5 ohms _______ d Neutral...

Страница 8: ...nd switch from the top to the EARTH CHASSIS CONNECTED left position 4 Using a calibrated resistance meter connect one lead to the GREEN banana jack on the LCT 601 Connect the other lead to exposed dea...

Страница 9: ...exposed surface of the EUT 10 Regarding the power switch of the EUT the test is to be conducted with the EUT power switch in both the ON and OFF positions 11 Read the current meter for all four condit...

Страница 10: ...ected lead is to be energized by properly switching the EUT 10 Connect all patient connected leads together connect them to the RED LEAKAGE banana plug of the LCT 601 and conduct measurements for all...

Страница 11: ...by properly switching the EUT 10 Connect all patient connected leads together connect them to the RED LEAKAGE banana plug of the LCT 601 and conduct measurements for all eight test conditions For thi...

Страница 12: ...jack on the LCT 601 to a patient connected lead 10 Connect the RED LEAKAGE banana plug of the LCT 601 to each patient connected lead in turn and conduct measurements for all four test conditions Each...

Страница 13: ...he RED and GREEN LEAKAGE banana plugs on the LCT 601 Set it to read 10uV 9 Connect the BLACK FLOATING MEAS POINT banana jack on the LCT 601 to a patient connected lead 10 Connect the RED LEAKAGE banan...

Страница 14: ...o voltage meter RED RESISTANCE jack N C FLOATING MEAS PT EUT nonisolated patient connections MEAS POINT Switch FLOATING EARTH Switch TEST IN BOTH POSITIONS NEUTRAL Switch CONNECTED POLARITY Switch TES...

Страница 15: ......

Страница 16: ...econd switch from the top to the EARTH CHASSIS CONNECTED left position 12 Using a calibrated resistance meter connect one lead to the GREEN banana jack on the LCT 601 Connect the other lead to exposed...

Страница 17: ...sed grounded metal on the EUT 9 Regarding the power switch of the EUT the test is to be conducted with the EUT power switch in both the ON and OFF positions 10 Read the current meter for both conditio...

Страница 18: ...h of the EUT the test is to be conducted with the EUT power switch in both the ON and OFF positions 10 Read the current meter for both conditions above A passing result is 300uA 300uV on the meter con...

Страница 19: ...EAKAGE banana plug of the LCT 601 and conduct measurements for both test conditions For this test switch the EUT to the output condition that results in the highest leakage output 10 Read the current...

Страница 20: ...patient connected lead is to be energized by properly switching the EUT 10 Read the current meter for the eight conditions noted above and for each patient connected output A passing result is 10uA 10...

Страница 21: ...ted lead in turn and conduct measurements for all four test conditions Each patient connected lead is to be energized by properly switching the EUT 11 Testing is done for each patient lead to all othe...

Страница 22: ...ed by properly switching the EUT 11 Testing is done for each patient lead to all other patient leads that is each pair combination must be tested 12 Read the current meter for both conditions noted ab...

Страница 23: ...ect a lead between the RED LEAKAGE banana plug on the LCT 601 and exposed grounded metal on the EUT 9 Regarding the power switch of the EUT the test is to be conducted with the EUT power switch in bot...

Страница 24: ...Phone 800 748 6224 Technical Assistance is available from Compliance West USA between the hours of 8 30 AM and 5 00 PM Pacific Time Compliance West USA 2120 Jimmy Durante Blvd Suite 124 Del Mar CA 92...

Страница 25: ......

Страница 26: ...capacitance meter measure the capacitance from 1 1 and verify 15 nF 5 Replace if outside this tolerance _______ 4 Using an ohmmeter measure the resistance from 2 2 and adjust the potentiometer to ach...

Отзывы: