14
1.
Shutdown Limit Setting: May be set too low. This would cause normal input capacitor
charging to draw more that the preset leakage current limit, triggering a Leakage Current
Fail light, and terminating the test. Consider raising the acceptable leakage current level;
see
Adjustment of the Leakage Current Shutdown point
.
If the Shutdown Limit level is at its highest setting, and failures continue, a DC hipot tester
may be required
2.
Ramp Time: May be set too fast. A very fast ramp time may allow input capacitors to act
as short circuits, triggering a Leakage Current Fail light and terminating the test. Consider
to have a slower Ramp Time, by increasing the time; see
Ramp Time Adjust.
Содержание HT-20KVPac
Страница 1: ...Dielectric Withstand Tester Model HT 20KVPac 0 20 000 Volts AC Output ...
Страница 2: ...i ...
Страница 4: ...iii ...
Страница 12: ...7 Figure 4 1 Front Panel Controls Indicators and Connectors ...