19
(50/60 Hz)
IEC61000-4-8
magnetic field shall
be at the level
characteristic of a
typical location in a
typical commercial
or
hospital
environment.
Note:
UT is the a.c. mains voltage prior to application of the test voltage.
Table 3: Electromagnetic immunity 2
Guidance and manufacturer's statement - Electromagnetic immunity
Comper SkinEnergy is intended for use in the electromagnetic environment as
specified below. The buyer or user of Comper SkinEnergy should assure that it is
used in such an environment:
Immunity
test
IEC
60601
Test
Level
Compliance
level
Electromagnetic
environment
-
Guidance
RF
conduction
IEC61000-4-
6
RF radiation
IEC61000-4-
3
3 V (r.m.s.)
150 kHz~80 MHz
3 V/m
80 MHz~2.5 GHz
Not
applicable
3 V/m
Portable
and
mobile
RF
communication equipment shall be
used not closer to any part of
Comper
SkinEnergy,
including
cables, than the recommended
separation distance. The distance
shall be calculated according to the
formula
applicable
to
the
frequency of the transmitter.
Recommended
separation
distance
d =
P
2
.
1
d =
P
2
.
1
80 MHz~800 MHz
d =
P
3
.
2
800 MHz~2.5 GHz
where:
P
The maximum rated output
power of the transmitter in watts
(W) according to the transmitter's
manufacturer;