6
The ThermArt thermometer is supplied with two AAA alkaline batteries. For best performance,
please insert new battery when the battery symbol “ ” appears on the display.
Precautions
1) Please install the forehead probe and ear probe cover correctly to ensure accurate measurement.
2) The forehead mode is contact measurement. Please contact the forehead probe to the center of
your forehead during measurement.
3) Please use clean and undamaged new ear probe cover for each ear temperature measurement.
4) When the probe needs to be cleaned, gently wipe the probe surface with 70% medical alcohol.
5) Please use the ear probe cover assigned by our company, or the measurement may be inaccurate.
Please contact
for assistance if you don’t kown where to purchase the ear
probe cover.
6) DO NOT move during measurement.
7) DO NOT force the probe into the ear to avoid ear canal injury.
8) The thermometer is intended for indoor use. Wind turbulence will adversely affect the accuracy
of the device. DO NOT take measurements in places with large airflow, such as facing to the fans
and air conditioner vent.
9) DO NOT use the device outdoors or under strong sunlight.
10) DO NOT hold the thermometer probes with your hand during measurement, so as to avoid the
influence of heat on the accuracy of the measurement results.
11) The measured part shall not be covered by hair. If you have sweat stains, wipe them with a dry
towel before taking a measurement.
12) If the patient feels pain or discomfort in the ear canal during ear temperature measurement,
please stop using this product immediately.
13) Make sure that both user and thermometer have stayed in steady-state room condition for at
least 30 minutes. Recent exposure to hot or cold temperatures will impact the reading.
14) If taking multiple measurements, it is advised to wait at least 5 seconds between measurements.
The thermometer is very sensitive to small changes in temperature, repeated use with no pause may
impact your reading.
15) DO NOT use the thermometer immediately after using ice pack or entering room from outside
in winter. In such cases, please wait until the forehead or ears are warm before taking a
measurement.
16) When the probe becomes dirty, gently wipe it with a soft dry cloth or cotton swab. Do not wipe
the probe with paper or towels.