Comfier B15BLU Скачать руководство пользователя страница 8

13

12

Patient Position:

1) Comfortably seated                       2) Legs uncrossed

3) Feet flat on the floor                      4) Back and arm supported

5) Middle of the CUFF at the level of the right atrium of the heart

NOTE: 

5.4. Irregular Heartbeat Detector     

5.5. Error Indicates 

Picture-06

The following symbol will appear on the display when measuring abnormal

Er 1

SYMBOL

No display 

appears

Weak battery or improper 

placement

Sensor abnormal 

Replace both batteries with new ones. 

Check the battery installation for proper 

placement of the battery polarities.

CAUSE

CORRECTION

Er 2

Er 3

Er 4

Er 5

Er 6

Er 7

Er 8

Check if the pump is working or not. If it 

is working, then the problem is sensor 

abnormal. Please send it to the local 

distributor.
Check if the air releasing is too slow or not. 

If it is too slow, please check if there is any 

dust in the tube plug of the cuff and the cuff 

port in the device. If yes, please clean and 

start the measurement again. If no, please 

send the device back to the local distributor.

Occasionally-measure for one more time/ 

Always - send it to local distributor

Tie the cuff correctly and make sure the air 

plug is properly inserted in the unit

Correct it and make the measurement again

Please keep quiet and don’t move

Please send back to the local distributor

Please send back to the local distributor

Monitor could not detect pulse 

wave or cannot calculate the 

blood pressure data

Measurement result is abnormal 

(SYS

35mmHg, DIA

23mmHg)

Too loose cuff or air leakage 

(Cannot inflate to 30mmHg within 

15s)

The air tube is crimped

The sensor is sensing great 

fluctuation in the pressure
The pressure that the sensor 

sensing is over the limit

The demarcation is incorrect or the 

device has not been demarcated

        This symbol - indicates that certain pulse irregularities were detected 

during the measurement.

In this case, the result may deviate from your normal basal blood pressure – 

repeat the measurement. 

Information for the 

physician on frequent appearance of the Irregular 

Heartbeat Symbol.

This instrument is an oscillometric blood pressure monitor device that also 

analyzes pulse frequency during measurement. The instrument is clinically 

tested.

If pulse irregularities occur during measurement, the irregular heartbeat 

symbol is displayed after the measurement. If the symbol appears more 

frequently (e.g. several times per week on measurements performed daily) 

or if it suddenly appears more often than usual, we recommend the patient to 

seek medical advice. The instrument does not replace a cardiac 

examination, but serves to detect pulse irregularities at an early stage.

Содержание B15BLU

Страница 1: ...8mm Fully Automatic Upper Arm Blood Pressure Monitor User Manual Model B15BLU COMFIER TECHNOLOGY CO LTD Address 573 BELLEVUE RD NEWARK DE 19713 USA www facebook com com ermassager supportus com er com...

Страница 2: ...ure and its Measurement 3 Components of Your Blood Pressure Monitor 4 Using Your Monitor for the First Time 5 Measurement Procedure 6 Care and Maintenance 7 Warranty Service 8 Certifications 9 Technic...

Страница 3: ...arning The separate ac adapter which is intended to connect USB interface of Blood Pressure Monitor has not been evaluated according to IEC 60601 1 The safety of the product shall be reappraised when...

Страница 4: ...s microwave ovens during use These can lead to erratic results Note Do not attempt to service or repair this device yourself Should a malfunction occur refer to local distributor or the manufacturer C...

Страница 5: ...y 4 Irregular heartbeat checking 6 Low battery display 8 Auto power off 10 Volume adjustment 1 Talking function 3 Cuff self checking function 5 Average value function 7 WHO function 9 External power a...

Страница 6: ...onmental pollution After you load the battery or connect power for the monitor long press the SET button for more than 3s and then you can start to set 4 2 System Settings Setting the User Press the M...

Страница 7: ...ortably in a chair with your feet flat on the floor 2 Select your User ID A or B 3 Stretch your arm forward on the desk and keep relaxing make sure the palm of hand is upturned Make sure arm is in cor...

Страница 8: ...ase keep quiet and don t move Please send back to the local distributor Please send back to the local distributor Monitor could not detect pulse wave or cannot calculate the blood pressure data Measur...

Страница 9: ...button can be pressed at any time The device then immediately lowers the cuff pressure automatically 5 7 Discontinuing a Measurement The following symbols may appear in this manual on the Digital Blo...

Страница 10: ...facturer s warranty g Batteries and electronic instruments must be disposed of in accordance with the locally applicable regulations not with domestic waste 6 1 Accuracy test Sensitive measuring devic...

Страница 11: ...nit This may result in incorrect operation of the unit Model B15S Weight 294 1g Batteries and AC adapter are not included Display 66 85mm 3 35 x2 6 LCD Digital Display Size 140 W x 120 L x 67 H mm 5 5...

Страница 12: ...d be at least 30 Mains power quality should be that of a typical commercial or hospital environment Mains power quality should be that of a typical commercial or hospital environment NOTE UT is the a...

Страница 13: ...ance d in metres m can be estimated using the equation applicable to the frequency of the transmitter where P is the maximum output power rating of the transmitter in watts W according to the transmit...

Отзывы: