background image

 
  

- 3 - 

PUTTING DEVICE INTO OPERATION 

1)

 

Mounting the connection adapter on the transducer 

 

unscrew  the  four  screws  at  the  corners  of  the 
transducer case and remove the cover 

 

if you have a new transducer, install it according to the 
transducer manual (keep it working from the supplied 
battery).  After  the  adapter  is  mounted,  the 
Configuration  button  inside  the  device  will  not  be 
accessible  -  therefore,  before  continuing,  make  sure 
that  the  device  is  turned  on  and  that  you  do  not 
require accelerated transducer parameter settings. 

 

remove  the  battery  by  pulling  on  its  handle  from  the 
ready transducer 

 

insert one SP015 connection adapter contact into the 
Wx8xx  transducer  battery  contact  field.  Then  insert 
the  second  contact  into  the  other  the  transducer 
battery contact field and close the cover. The contacts 
of the two devices must fit exactly together. 

 
 

 

verify  that  the  connection  adapter  is  flush  with  the 
transducer housing circuit. If not, remove the adapter 
and  repeat  the  assembly  procedure.  Under  no 
circumstances should you attempt to force the adapter 
and transducer into the correct position. 

 
 
  
 

 

if  the  adapter  is  in  the  correct  position,  screw  it  with 
the original screws 

 

keep  the  original  cover  and  battery  for  case  the 
adapter is will removed 

 

 

 

 