IE-LOG-S0841-07 

 

  3 

Minimum pulse duration at binary input: 500 ms (shorter pulses may not be recorded) 
Maximum frequency at binary input: 0.5 Hz (i.e. maximum 5 pulses in 10 s) 
Power current of connected contact: 3  A (contact closed) 
Voltage across open contact: maximum 3.6 V 
Cable parameters for connection of the binary input: 2 wires with common shielding 

(detailed information on cable – see page 8) 

Cable length for connection of the binary input: maximum 10 m 
Measurement interval measurement, alarm evaluation, memory MIN/MAX update and 

display refresh: 10 s 

Logging interval to memory: 10 s to 24 h (20 steps) 
Memory capacity:  

for non-cyclic logging mode 

16 252 records for each temperature channel 

for cyclic logging mode 

 

15 296 records for each temperature channel 

  Specified values are maximum possible and can be reached only if record since last 

memory erasing was not interrupted and no event on binary inputs was recorded. Each 
event record on the binary input decreases above capacity of one record.  

Communication with computer: via RS232 (serial port) by means of COM adapter or USB 

port by means of USB adapter; data transfer from logger via communication adapter is 
optical 

Real time clock: adjustable from computer, integrated calendar including leap years 
Error of internal RTC: < 200 ppm (i.e. 0.02 %, 17.28 s in 24 h) 
Power: Lithium battery 3.6 V size AA 
Battery life: 

typical (data download to PC is carried out about once a week): 5 years 
in continuous on-line mode with interval 1 minute: approximately 3 years 
in continuous on-line mode with interval 10 seconds: approximately 1 year 
Notice: The above battery lives are valid for logger operation in temperature from -5 to 
+35°C. If often operated outside of the above temperature range battery life can be reduced 
to 75%. 

Protection: IP67 
Operational conditions: 
  Operational temperature range: -30 to +70 °C 
  Operational humidity range: 0 to 100 %RH 
   Specification of outer characteristics accordingly to Czech National Standard 33-2000-3: 

normal environment accordingly to appendix NM:  AE1, AN1, AR1, BE1 

   Operational position: negligible  
  Logger installation: by self adhesive Dual Lock, applied to clean, flat surface 

Not allowed manipulation: it is not allowed to operate instrument with not connected 

temperature probes - if all probes are not connected, unconnected connectors should be 
protected by included connector cover. 

Limit condition: temperature -40 to +70 °C, humidity 0 to 100 %RH 
Storing condition: temperature -40 to +85 °C, humidity 0 to 100 %RH 
Dimensions: 93 x 64 x 29 mm 
Weight including battery: approximately 120 g 
Material of the case: ABS 
 
 

 

 

Содержание S0841

Страница 1: ...LOGGER S0841 Dual channel temperature logger with two binary inputs and LCD display Instruction Manual ...

Страница 2: ...e when alarm is indicated permanently till alarm memory reset Alarm function is enabled or disabled for each channel individually Binary channels have no alarm modes Logging mode can be adjusted as non cyclic when logging stops after filling the memory In cyclic mode oldest stored values are overwritten by new In addition logging mode can be selected when logging is active only if measured value i...

Страница 3: ...lock adjustable from computer integrated calendar including leap years Error of internal RTC 200 ppm i e 0 02 17 28 s in 24 h Power Lithium battery 3 6 V size AA Battery life typical data download to PC is carried out about once a week 5 years in continuous on line mode with interval 1 minute approximately 3 years in continuous on line mode with interval 10 seconds approximately 1 year Notice The ...

Страница 4: ...heck or optionally set the logger real time clock select suitable logging interval it has no influence to binary inputs record select logging mode cyclic or non cyclic set for each channel type of temperature probe Pt1000 or Ni1000 disable channels for logging with not connected temperature probe switch ON the logger or switch OFF if it is about to be switched ON by magnet or automatically with de...

Страница 5: ... permanently connected to the computer using of magnet start stop is disabled To enable the logger control by magnet is suitable only in cases when possibility of unauthorized manipulation to the logger operation is eliminated Reading on the display in usual operation logger switched ON After switching ON the logger all LCD symbols are displayed for checking the display Then actual date and time i...

Страница 6: ...ing therefore does not run Displayed symbol indicates logger is ON If information on actual measured values is needed it is possible anytime to display reading display by means of the magnet only if communication adapter is not connected permanently Plug magnet into guide slots from logger front side for approximately 4 s and wait till reading on the display appears f logger has enabled the functi...

Страница 7: ... or battery disconnection for longer period than approximately 30 s after display switch ON during date and time display all four arrows appears as warning to check or set it again from the computer However all logger functions work without limitation If reading BAT is displayed periodically on LCD upper line for 1 s with 10 s interval the end of estimated battery life is coming however logger fun...

Страница 8: ...ry to enable the function from the PC software before Notice It is not possible to combine this function with function of switching logger OFF by magnet User software enables to select only one of them or none Plug magnet to guide slots from logger front side and wait approximately 1 s for decimal point appears right on LCD upper line After decimal point appearance it is necessary immediately till...

Страница 9: ...ery replacement This step is necessary to evaluate properly the remaining battery capacity Old battery or logger itself after its life is necessary to liquidate ecologically Connection of female ELKA 3008V connector for temperature probe and binary contact It is possible to order as an optional accessory connector for connection of temperature probe Signal cable should be shielded round cross sect...

Отзывы: