Replacement procedure: 

 

switch off the logger by the PC program or by magnet (if low battery allows) 

 

unscrew four corner screws and remove the lid 

 

remove old battery by pulling the glued tape 

 

insert new battery 

respecting the correct polarity

 (see s and – near battery holder). If 

you  connect  new  battery  up  to  30  s,  all  logger  settings  remain unchanged. In the opposite case 
check by means of PC program all settings, especially the real time clock in logger. 

Attention, 

inserted battery with incorrect polarity causes logger damage!

 

 

put the lid back again and screw four screws. Be careful to have the rubber sealing properly in the 
slot and tighten the screws properly to ensure the water resistance of the instrument. 

connect  logger  to  the  computer  and 

write  to  it  the  information  on  battery  replacement 

(menu Configuration/Battery replacement). This step is necessary to evaluate properly the remaining 
battery capacity. 

 
 
 

Old battery or logger itself (after its life) is necessary to liquidate ecologically!

 

 
 
 
 
 
 
 

Connection of female ELKA 3008V connector for temperature probe and voltage contact 

It  is  possible  to  order  as  an  optional  accessory  connector  for 

connection of temperature probe. Signal cable should be shielded, round 
cross section of minimum diameter of 3.5 mm and maximum diameter of 
5 mm, maximum length of 10 m and as large as possible cross section of 
cords  (stranded  wires).  Outer  diameter  must  be  from  1  to  1.55  mm 
including  insulation.  Connect  connector  to  temperature  signal 
accordingly  to  the  figure  by  means  of  the  manual  included  in  delivery. 
Choose  the  most  suitable  plastic  insert  accordingly  to  the  diameter  of 
stranded  wires  of  the  cable.  The  IP67  protection  is  guaranteed  only  if 
connector is assembled correctly! 

It  is  necessary  to  select  and  connect  cable  for  voltage  inputs  with 

respecting the same rules as above. Connect the voltage signals instead of the RTD probe. Connect 
positive pole to the brown terminal and signal ground to the blue terminal. Due to high impedance 
of  voltage  inputs  it  is  necessary  to  avoid  places  with  higher  electromagnetic  interference  when 
installing cables from contacts (switchboards, motors, contactors, frequency converters, etc.). 

 
 

Instrument passed through tests of electromagnetic compatibility (EMC): 

 

Device conforms in accordance with EN 61326-1 these norms: 

 

radiation:  

EN 55011 Class B 

 

immunity:  

EN 61000-4-2 (levels 4/8 kV, Class A) 

 

 

 

EN 61000-4-3 (intensity of electric field 3 V/m, Class A) 

 

 

 

EN 61000-4-4 (levels 1/0.5 kV, Class A) 

 

 

 

EN 61000-4-6 (intensity of electric field 3 V/m, Class A) 

 

Содержание S0541

Страница 1: ...LOGGER S0541 Dual channel temperature logger with two voltage inputs 0 5V Instruction Manual...

Страница 2: ...d or disabled for each channel individually Logging mode can be adjusted as non cyclic when logging stops after filling the memory In cyclic mode oldest stored values are overwritten by new In additio...

Страница 3: ...75 Protection IP67 Operational conditions Operational temperature range 30 to 70 C Operational humidity range 0 to 100 RH Specification of outer characteristics accordingly to Czech National Standard...

Страница 4: ...record runs permanently or only if alarm is active If alarms are about to be applied set both limits for each measured quantity and enable alarm optionally enable permanent alarm indication alarm wit...

Страница 5: ...larity is not allowed increased consumption from the battery Note all GND terminals are galvanically connected in data logger 1 Legend of connection of connector ELKA INx brown input GND blue ground G...

Страница 6: ...her than 90 If display is OFF and logger is in mode when record runs only when alarm is active the LOG symbol is replaced by adjacent symbol hyphen It appears in case all measured values are inside of...

Страница 7: ...New initialization of logger can occur in switching ON the logger immediately after displaying of all LCD segments for checking e g after replacement of totally discharged battery for new one State i...

Страница 8: ...ing logger OFF by magnet User software enables to select only one of them or none Plug magnet to guide slots from logger front side and wait approximately 1 s for decimal point appears right on LCD up...

Страница 9: ...nection of temperature probe Signal cable should be shielded round cross section of minimum diameter of 3 5 mm and maximum diameter of 5 mm maximum length of 10 m and as large as possible cross sectio...

Отзывы: