Comet S0541 Скачать руководство пользователя страница 8

 

  IE-LOG-S0541-03 

 

Switching the logger ON by magnet 

Plug magnet to guide slots from logger front side and wait approximately 1 s for decimal point 

appears right on LCD upper line. After appearance it is necessary immediately (till indication point 
is displayed) to remove magnet from guide slots and logger switches ON. 

 

 

 

Switching the logger OFF by magnet 

The  procedure  is  identical  with  the  above  procedure  for  switching  ON.  If  decimal  point  does 

not appear after 1 s, it is necessary remove the magnet and repeat the procedure. 
 

Reset of MIN/MAX values by magnet 

Function enables to clear MIN/MAX values by magnet without using computer. It is necessary 

to enable the function from the PC software before. 

Notice: 

It  is  not  possible  to  combine  this  function  with  function  of  switching  logger  OFF  by 

magnet! User software enables to select only one of them (or none). 

Plug magnet to guide slots from logger front side and wait approximately 1 s for decimal point 

appears  right  on  LCD  upper  line.  After  decimal  point  appearance  it  is  necessary  immediately  (till 
indication point is displayed) to remove magnet from guide slots. Reading CLR MIN MAX appears 
for several seconds and MIN/MAX values will be cleared. 

 

 

 

Battery replacement 

Low  battery  is  indicated  on  the  display  by  blinking  of  reading  "BAT".  It  can  be  displayed 

permanently,  if  battery  voltage  is  too  low.  Replace  the  battery  for  new  one.  If  logger  is  operated 
often  in  temperature  below  -5°C  or  over  +35°C  and  PC  program  indicates  remaining  battery 
capacity below 25% it is also recommended to replace the battery. Applied is Lithium battery 3.6 V, 
size AA. Battery is located under logger lid. 

Warning:

  near  battery  fragile  glass  reed  contact  is  located  –  be  careful  not  to  damage  it.  Be 

careful in battery replacement! 

 
 
 

Содержание S0541

Страница 1: ...LOGGER S0541 Dual channel temperature logger with two voltage inputs 0 5V Instruction Manual...

Страница 2: ...d or disabled for each channel individually Logging mode can be adjusted as non cyclic when logging stops after filling the memory In cyclic mode oldest stored values are overwritten by new In additio...

Страница 3: ...75 Protection IP67 Operational conditions Operational temperature range 30 to 70 C Operational humidity range 0 to 100 RH Specification of outer characteristics accordingly to Czech National Standard...

Страница 4: ...record runs permanently or only if alarm is active If alarms are about to be applied set both limits for each measured quantity and enable alarm optionally enable permanent alarm indication alarm wit...

Страница 5: ...larity is not allowed increased consumption from the battery Note all GND terminals are galvanically connected in data logger 1 Legend of connection of connector ELKA INx brown input GND blue ground G...

Страница 6: ...her than 90 If display is OFF and logger is in mode when record runs only when alarm is active the LOG symbol is replaced by adjacent symbol hyphen It appears in case all measured values are inside of...

Страница 7: ...New initialization of logger can occur in switching ON the logger immediately after displaying of all LCD segments for checking e g after replacement of totally discharged battery for new one State i...

Страница 8: ...ing logger OFF by magnet User software enables to select only one of them or none Plug magnet to guide slots from logger front side and wait approximately 1 s for decimal point appears right on LCD up...

Страница 9: ...nection of temperature probe Signal cable should be shielded round cross section of minimum diameter of 3 5 mm and maximum diameter of 5 mm maximum length of 10 m and as large as possible cross sectio...

Отзывы: