Press HOLD  key in the default mode (displaying of actual measured value) to store actual 

measured values to internal memory (indicated by short beep). Anytime it is possible to display 

stored values from MENU (see below). Each pressing of the HOLD key in the default mode causes  

values stored in HOLD memory are replaced with actual ones. 

 

Switched ON instrument permanently updates minimum and maximum 

memory of each measured values. Press MIN key (resp. MAX key) in the 

default mode to display minimum (resp. maximum) reading. These 

minimum and maximum readings are indicated by MIN (MAX) symbols on 

the LCD. Press DISPLAY key to display minimum (resp. maximum) value 

of other values. Pressing MIN (MAX) or MENU key again to return to 

default mode. Minimum and maximum memory is cleared from menu after 

confirmation selection CLR (see below). Values in HOLD, MIN and MAX 

memories remain stored even after instrument is switched OFF. 

 
 

Functions and settings available from menu 

 

Press MENU key to enter mode of viewing menu items one by one. Press arrow keys up and 

down to list all menu items. Press MENU key again to return to default mode (displaying of actual 

measured values). 

  

 

 
Press ENTER key to enter alarm setting mode (see below). 

 

 
 
 

This item indicates if audio signaling of alarm indication is switched on 
(On) or switched off (OFF). Press ENTER key to change actual setting. 
Notice: if the battery voltage is low, audio indication is out of operation to 
reduce current consumption independently on this selection. 
 
 
Clearing of minimum and maximum memory of all values. Memory is 

cleared after pressing ENTER key. Clearing is confirmed by reading YES 

on the LCD lower display. 
 

 
 

Press ENTER key to display values stored in the HOLD memory. Press 

DISPLAY key to display other stored values. Press MENU key to leave this 

mode. 
 
 
 

 

  IE-COM-C0321-04 

Содержание Commeter C0321

Страница 1: ...COMMETER C0321 Dual channel thermometer for thermocouples J K S Instruction manual...

Страница 2: ...0 7 s in FAST mode 0 7 to 5 s in dynamic mode Power battery 9V or ac dc adapter 12V with NiCd accumulator 9V Average current consumption 0 3 to 1 3 mA depending on operation mode Protection IP20 Opera...

Страница 3: ...on instrument then starts the measurement mode and actual measured values are displayed If instrument is ON do not disconnect or connect the probes to prevent storing of incorrect value to minimum and...

Страница 4: ...tion CLR see below Values in HOLD MIN and MAX memories remain stored even after instrument is switched OFF Functions and settings available from menu Press MENU key to enter mode of viewing menu items...

Страница 5: ...battery life significantly Select the desired mode by ENTER key Notice if battery voltage is low the FAST mode is out of operation to reduce current consumption independently on this selection Pressi...

Страница 6: ...t setting Alarm activation of each value can be disabled by setting lower alarm limit of the desired value up to its maximum This is indicated by OFF reading at the position of numeric value Value of...

Страница 7: ...propriate menu selection to set again LCD refreshment mode d REF alarm audio indication AUDI and clear the minimum and maximum memory CLR If the instrument does not work after replacement of totally d...

Отзывы: