Comet Commeter C0321 Скачать руководство пользователя страница 2

Manual for use of dual channel thermometer for thermocouples J, K, S 

COMMETER C0321 

 

Instrument is designed for measurement of temperature from two thermocouples of types J, K, S 
with the possibility to display the temperature difference between channels. Measured values are 
displayed on a dual line LCD display. Instrument compares measured values of temperature with 
two adjustable levels for each measured channel. Breaking the level is indicated by blinking the 
proper value on display and by audio indication (switchable). Instrument is equipped with minimum 
and maximum memory and Hold function. Minimum and maximum values and Hold value are 
possible to display on the LCD anytime. 

 
Technical parameters: 
 

Parameters of measurement (both channels): 

 

Range of measurement (with the limitation of connected probe parameters: 

 

 

thermocouple J (Fe-Co) 

 

-200 to +750°C 

 

 

thermocouple K (NiCr-Ni) 

 

-200 to +1300°C 

 

 

thermocouple S (Pt10%Rh-Pt) 

  -50 to +1700°C 

 Resolution: 
 

 

thermocouples J and K        0.1°C from –99.9 to +999.9°C, otherwise 1°C 

 

 

thermocouple S 

         0.4°C from –50.0 to +999.9°C, otherwise 1°C 

 

Accuracy of inputs without probes: 

thermocouples J and K 

±(0.1% +0.4°C) from reading 

thermocouple S 

 

±(0.1% +0.85°C) from reading 

 

 

      

Compatible temperature probes: J,K,S thermocouple with subminiature male thermocouple 
connector 

 

Compensation of thermocouple cold junction temperature: automatic from –10 to +60°C 

Measuring interval and display reading refresh: approximately 0.7 s in FAST mode 

 

 

 

 

 

 

        

           0.7 to 5 s in dynamic mode 

 

Power: battery 9V or ac/dc adapter 12V with NiCd accumulator 9V 

 

Average current consumption: 0.3 to 1.3 mA (depending on operation mode) 

 Protection: 

IP20 

 Operation 

conditions: 

 

  Ambient temperature range: -10 to +60 °C 

 

  Ambient relative humidity range: 5 to 95 % RH, non condensing 

  Outer characteristics in accordance with EN 33-2000-3: normal environment with   
characteristics AD1, AE1, AF1, AG1, AH1, AK1, AL1, AN1, AP1, AQ1, AR1, AS1, BA1, 
BE1 

 

Storing conditions: temperature -10 to +60 °C relative humidity 5 to 95 % RH 

 

Dimensions: 141 x 71 x 27 mm 

 

Weight including battery: approximately 150 g 

  

 

Battery life depends on selected display refresh mode (see below). In FAST mode display is 
refreshed in shortest possible interval with highest current consumption. In dynamic mode display 
is refreshed in interval up to 5 s in case measured values remain stable. Refresh interval is shortened 
to approximately 0.7 s only if measured values change. Current consumption in this mode in usual 
operation is lower, battery life is up to 4 times longer. The FAST mode is recommend to use only in 
cases, when slower display response is not acceptable. 
Battery voltage drop below 7 V is indicated with blinking of "BAT" in default display mode 
(displaying of actual values) and FAST mode is automatically cancelled to save the battery. At the 
same time audio indication of alarms is automatically switched OFF. 
If instrument is supplied from external ac/dc adapter, internal 9V battery should be replaced with 
rechargeable NiCd accumulator. In usual operation from adapter accumulator is charged only with 

 

  IE-COM-C0321-04 

Содержание Commeter C0321

Страница 1: ...COMMETER C0321 Dual channel thermometer for thermocouples J K S Instruction manual...

Страница 2: ...0 7 s in FAST mode 0 7 to 5 s in dynamic mode Power battery 9V or ac dc adapter 12V with NiCd accumulator 9V Average current consumption 0 3 to 1 3 mA depending on operation mode Protection IP20 Opera...

Страница 3: ...on instrument then starts the measurement mode and actual measured values are displayed If instrument is ON do not disconnect or connect the probes to prevent storing of incorrect value to minimum and...

Страница 4: ...tion CLR see below Values in HOLD MIN and MAX memories remain stored even after instrument is switched OFF Functions and settings available from menu Press MENU key to enter mode of viewing menu items...

Страница 5: ...battery life significantly Select the desired mode by ENTER key Notice if battery voltage is low the FAST mode is out of operation to reduce current consumption independently on this selection Pressi...

Страница 6: ...t setting Alarm activation of each value can be disabled by setting lower alarm limit of the desired value up to its maximum This is indicated by OFF reading at the position of numeric value Value of...

Страница 7: ...propriate menu selection to set again LCD refreshment mode d REF alarm audio indication AUDI and clear the minimum and maximum memory CLR If the instrument does not work after replacement of totally d...

Отзывы: