IE-SNC-Tx6xx-07
43
Computed quantities
Dew point temperature
Range:
-
60 to +80°C (
-20 to
176°F
)
Accuracy:
±1
.
5°C (±2
.
7°F)
at ambient temperature
T < 25°C (77°F) a RH >
30%, for more details see graphs:
Specific humidity
This value depends on atmospheric pressure. Pressure for quantities calculation is stored in
device memory. Default value is 1013hPa and can be changed by software.
Range:
0 to 550g/kg
Accuracy:
±
2.1g/kg at ambient temperature T < 3
5°C (
95
°F)
Mixing ratio
This value depends on atmospheric pressure. Pressure for quantities calculation is stored
in device memory. Default value is 1013hPa and can be changed by software.
Range:
0 to 995g/kg
Accuracy:
±
2.2g/kg at ambient temperature T < 3
5°C (
95
°F)