Comet System COMMETER C3121P Скачать руководство пользователя страница 2

 2

 

 

IE-COM-C3121P-1

 

Manual for use of thermometer-hygrometer COMMETER C3121P 

 

Instrument is designed for measurement of temperature and relative humidity with an external 

probe  on  a  cable  with  the  possibility  of  displaying  the  dew  point 
temperature.  Measured  values  are  displayed  on  a  dual  line  LCD 
display.  Temperature is  measured by RTD sensor Ni1000/6180ppm. 
Instrument  compares  measured  values  of  temperature,  humidity  and 
dew  point  with  two  adjustable  levels  for  each  measured  quantity. 
Breaking  the  level  is  indicated  by    blinking  the  proper  value  on 
display and by audio indication (switchable). Instrument is equipped 
with minimum and maximum memory and Hold function. Minimum 
and  maximum  values  and  Hold  value  are  possible  to  display  on  the 
LCD anytime. 

The  sensing  element  of  the  device  is  an  external  probe  (see 

picture) designed for compressed air measurement up to 25 bars. The 
probe  has  a  cable  (length  of  1 to 4  meters)  with  connector. 
Mechanical  connection  of  the  probe  is  threaded  G1/2  with  O-ring 
seal. . It is good to mount probe directly into high pressure area (i.e. 
air  pressured  pipe)  if  it  is  possible.  There  is  other  possibility  to 
use flow chamber SH-PP, see Appendix A. 

The probe is an noninterchangeable part of the device. 

 

Before you remove probe, make sure that the pressure in a pressure chamber (duct, pipe…) 
and the ambient pressure are in equilibrium! 

 

Read instruction manual before the first device connection, please. 

 

Technical parameters: 

Parameters of measurement: 

Temperature: Range of measurement: -30 to +105 °C 

 

 

 

Resolution: 0.1 °C 

 

 

 

Accuracy: ±0.4 °C 

Humidity: 

Range measurement: 0 to 100 %RH 

 

 

 

Resolution: 0.1 %RH 

 

 

 

Accuracy: ±2.5 % RH at the range of 5 to 95 %RH 

Measuring temperature and humidity range is limited in accordance with the graph. 

  
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

  

Содержание COMMETER C3121P

Страница 1: ...COMMETER C3121P Digital thermometer hygrometer with external probe for compressed measurement Instruction manual...

Страница 2: ...element of the device is an external probe see picture designed for compressed air measurement up to 25 bars The probe has a cable length of 1 to 4 meters with connector Mechanical connection of the...

Страница 3: ...ge current consumption 0 15 to 0 7 mA depending on operation mode Protection device IP20 sensors are located in cover with IP40 protection Typical life of Zinc Chloride battery 4 months Typical life o...

Страница 4: ...ator In usual operation from adapter accumulator is charged only with small current If accumulator is totally discharged its full charging in instrument takes approximately 100 hours Instrument with a...

Страница 5: ...ault mode Minimum and maximum memory is cleared from menu after confirmation selection CLR see below Values in HOLD MIN and MAX memories remain stored even after instrument is switched OFF Functions a...

Страница 6: ...ication can be activated if menu AUDI On is selected see setting described above Alarm activation of each value can be disabled by setting lower alarm limit of the desired value up to its maximum This...

Страница 7: ...ions and data in memory HOLD MIN and MAX will be lost For the same reason do not disconnect the battery for longer than 1 minute even if instrument is switched OFF If it happens or if battery is total...

Страница 8: ...rupting the main line closing valve 5 the sample flow is regulated by this valve Measurement accuracy is typically not affected by the sample flow rate but at low speeds increasing response time outle...

Отзывы: