6.0

 

  LISN

 

T

HEORY

 

 

Line  Impedance  Stabilization  Networks  provide  standardized  line  impedance  to  the  EUT 
during  conducted  emissions  testing  which  is  independent  of  the  external  power  line 
impedance.  The  standardized  impedance  enables  consistent  readings  for  RF  noise 
measurements  on  the  power  line.  The  model  LI-150C  consists  of  two  separate  LISN 
networks  housed  in  separate  enclosures  for  testing  EUTs  with  single  phase  power 
requirements.  
 
In addition to providing standardized impedance, the LISN also acts as a low pass filter for 
the  power  to  equipment  under  test.  The  LISN  blocks  RF  noise  from  the  power  line  from 
reaching  the  equipment  under  test,  however,  the  power  to  the  equipment  passes 
through  the  LISN  with  minimal  attenuation.    This  filter  is  comprised  of  a  single  stage  low 
pass  LC  filter.  The  inductors  (L)  used  in  the  LI-150C  are  air  core  type  to  eliminate  the 
possibility of saturation and to provide stability.   
 
The  LISN  also  provides  a  low  impedance  path  for  the  RF  noise  from  the  EUT  to  the 
measuring  equipment,  thereby  facilitating  the  measurement  of  the  EUT  RF  noise.    The 
insertion  loss  factors  provided  with  every  LISN  network.  It  may  be  significant  at  low 
frequencies;  especially  below  400  kHz.  Therefore,  the  insertion  loss  correction  factors 
should always be considered for the highest measurement accuracy. 
 
Example  schematic  of  the  LISN  as  well  as  the  respective  impedance  requirements  are 
specified  in  the  ANSI  C63.4  and  CISPR-16-1-2  standards.  CISPR  16-1-2  also  specifies 
additional requirements for phase performance and Mains port isolation.  A typical circuit 
diagram of an LISN is given in figure below. 

 

 

Figure 5: Example of equivalent LISN circuit 

 
 
 
 

Содержание LI-150C

Страница 1: ...MANUAL LI 150C LINE IMPEDANCE STABILIZATION NETWORK 19121 El Toro Rd Silverado California 92676 949 459 9600 com power com REV021317 INSTRUCTION MANUAL For LINE IMPEDANCE STABILIZATION NETWORK Model L...

Страница 2: ...cription 10 5 1 Overview 10 5 2 Product Setup 10 5 3 Front Rear Panel Marking 11 6 0 LISN Theory 13 7 0 Typical Performance Data 14 8 0 Warranty 16 8 1 Maintenance 16 List of Figures Figure 1 Single P...

Страница 3: ...front and rear panel ports controls and indicators product specifications safety precautions operational instructions and warranty information and guidelines and instructions for its proper usage Info...

Страница 4: ...Networks CDN Comb Generators Current Probes Emissions Test Systems Conducted Immunity Test Systems Impedance Stabilization Networks ISN Line Impedance Stabilization Networks LISN Antenna Masts Near Fi...

Страница 5: ...e product The yellow triangle with a lightning bolt indicates an alert to the user that un insulated dangerous voltages are present within the product enclosure and on output connectors These voltages...

Страница 6: ...items ISO 17025 calibration data and certificate External transient limiter model LIT 153 3 4 General Safety Instructions The following safety instructions have been included in compliance with safet...

Страница 7: ...under EN 61010 1 standard safety requirements for electrical equipment for measurement control and laboratory use In addition the basic insulation required for a category I protection device cannot be...

Страница 8: ...b upon the instrument or use it as a step or ladder ENVIRONMENTAL CONDITIONS This equipment is designed for indoor use Ambient temperature range during operation should be between 5 C to 40 C STORAGE...

Страница 9: ...6 1 2 Edition 2 0 2014 03 ANSI C63 4 2014 Number of lines Two lines Max current rating 50 Amp AC 35 Amp DC Max Voltage rating 270 VAC 50 60 Hz Line to Ground 380 VDC Inductor type 50 H EUT Power conne...

Страница 10: ...eby minimizing its influence on the measurements and couples the disturbance voltages to the coaxial measurement port for connection to the measuring instrument like EMI Receiver or Spectrum Analyzer...

Страница 11: ...arth leakage current the device cannot be connected to any power source protected by either a ground fault circuit interrupter GFCI circuit breaker or residual current devices RCDs The power source mu...

Страница 12: ...3 4 EMI measurement Port The LI 150C provides a 50 N Type female connector to connecting the Spectrum analyzer or EMI receiver for making conducted noise measurements The shielding connection of the...

Страница 13: ...wever the power to the equipment passes through the LISN with minimal attenuation This filter is comprised of a single stage low pass LC filter The inductors L used in the LI 150C are air core type to...

Страница 14: ...pared to ANSI C63 5 CISPR 16 1 2 requirements Figure 7 Typical Phase compared to CISPR 16 1 2 requirements 20 30 40 50 60 70 0 1 1 10 100 Impedance ohms Frequency MHz LISN Impedance CISPR 16 1 2 Ed 1...

Страница 15: ...om power com REV021317 Figure 8 Typical Isolation compared to CISPR 16 1 2 requirements Figure 9 Typical Insertion Loss 20 30 40 50 60 70 80 90 100 0 1 1 10 100 Isolation dB Frequency MHz LISN Isolati...

Страница 16: ...directly or indirectly from the use or handling of this product or any inability to use this product separately or in combination with any other equipment When requesting warranty services it is recom...

Отзывы: