5.3.6 Fuse
The fuse is for the Fan over current protection, it is a 20 x 5mm 1A, 250V (T)
fuse. For continued protection against risk of fire, replace only with fuse of
the specified type, current and voltage ratings
5.3.7 Fan switch
This switch controls the Fan operation. Always turn the fan ON before
powering the EUT via LISN.
CAUTION
The LI-1100 LISNs uses forced air system to keep the air core inductors cool
during the test. Always use the supplied DC adapter to power the internal
fan during the test. Do not obstruct the airflow vents. Use the LISNs in a
ventilated area. Failure to observe these precautions may damage the
LISN. Overheated LISN may also provide incorrect measurement data. In
addition the heat may start a fire or injure test personnel. If the fan stops
working discontinue using the LISN and contact Com-Power for further
assistance.