![Cole Parmer pH 550 Скачать руководство пользователя страница 4](http://html1.mh-extra.com/html/cole-parmer/ph-550/ph-550_operation-manual_2635379004.webp)
4
1.
Be sure to remove the power before attempting to clean the
meter.
2.
Use a lint free cloth and clean water or neutral detergent.
3.
Wipe the outer surface of the instrument only.
4. Wipe dry the instrument before power again.
II. GENERAL INTRODUCTION
The Cole-Parmer pH 550 (
p
H and
T
emperature) System is a rugged
microprocessor based instrument assembled in a watertight ¼ DIN
case, designed for use in laboratories and process control
applications.
The pH 550 microprocessor allows the user to easily recalibrate the
parameters for the probes. The microprocessor also performs a self-
diagnostic routine every time you turn on the unit providing you
with basic information about the stability of the instrument.
The system simultaneously displays pH, Temperature, Alarm status
and current output in one LCD graphic screen. The LCD also
includes a backlight for dark environments. This system uses glass
electrode for the pH and a precise thermistor for temperature
providing you with accurate readings for all your measurements.
The pH 550 is equipped with three programmable high or low
relays for pH and one programmable high or low relay for
temperature; all relays are hysteresis driven. The system also has
two
isolated 4-20mA
analog output, offset and span configurable
for the pH & temperature display.
The pH 550 comes with a
RS485
interface that can easily let the user
log all data with an IBM
©
PC/AT compatible computer.