plot the output. (See chapter
VI
.
4-20 mA OUTPUT
.)
3.
TRANSMITTER OUTPUT
– This option would let you choose if
the current output type is linear or antilog.
The current output detailed see section
C. pH & TEMP. LINEAR
CURRENT OUTPUT
on page 20
.
f. TEMP.
CURRENT SETTING
page
1.
4mA OUT
– This value will be
used in conjunction with 20 mA
to plot the current output. (See
chapter
VI
.
4-20 mA OUTPUT
.)
2
. 20mA OUT
– This value will
be used in conjunction with the
4 mA value to plot the output.
(See chapter
VI
.
4-20 mA
OUTPUT
.)
The current output detailed see section
C. pH & TEMP. LINEAR
CURRENT OUTPUT
on page 20
.
g.
TEMP. CONTROL SETTING
1.
RELAY 4
- the temperature
has only one relay to control you
need to set what action it will
use, HIGH or LOW action. . (In
HIGH-action the relay will turn
ON
if the temperature is greater
or equal to RELAY4 set point, in
LOW-action the relay will turn
OFF
if the temperature is less
than or equal to RELAY4 set
point, which is modified by the hysteresis value (See chapter
V.
CONTROLLING THE RELAYS
.)