background image

3

General Description

The Wax Dispenser is designed for on-demand delivery of molten wax. The novel design of the instrument ensures
that it takes up a minimum bench space in the laboratory, whilst still maintaining a large 7.5L capacity tank, for
melting up to 6kg of pelletized wax at a time. The Wax Dispenser has digital temperature control which provides
accurate temperature control of the molten wax, and is coupled with an ultra fast heating system for rapid melting
of pelletized wax. Delivery of molten wax is via a non-drip lever tap, which is heated via a dedicated heating
system, which prevents blockages resulting from solidified wax and ensures an even flow of wax. The inner tank has
a filter screen (0.5mm mesh) fitted to prevent coarse particles from blocking the delivery tap. The tank is also fully
insulated to prevent heat loss from the tank and to ensure that the outer surfaces of the tank are safe to touch.

Important Safety Advice

This product is designed for laboratory use only. Always follow good laboratory practice.
Users should be aware of the following safety advice:

v

 

DO NOT

 use without appropriate training.

v

 

SHOCK HAZARDS OR UNSAFE PRACTICES ARE DANGEROUS

 as they can cause severe personal injury, fire  

 

or death.

v

 

DO NOT

 use combustible substances near hot objects.

v

 

DO NOT

 use the equipment in hazardous atmospheres.

v

 

DO NOT

 immerse the equipment in water.

v

 

DO NOT

 operate or handle any part of the product with wet hands or use on surfaces that may become flooded.

v

 

NEVER

 move the product while still connected to the power supply.

v

 

HIGH TEMPERATURES ARE DANGEROUS

 as they can cause serious burns to operators and ignite combustible 

 material.

v

 

USE CARE AND WEAR PROTECTIVE GLOVES TO PROTECT HANDS

.

v

 

NEVER

 lift or carry the instrument during operation.

v

 

NEVER

 use the tap to lift or carry the instrument.

DO NOT

 position the unit so that it is difficult to disconnect from the mains supply using the mains plug.

The mains outlet socket used should be located close to the equipment and readily identifiable and accessible

 

to users.

v

 

DO NOT

 leave equipment switched on and it is not recommended to leave any heating apparatus unattended 

 

during operation.

v

  The unit should be carried using both hands.

v

 

DO NOT

 allow molten wax to accumulate on the surface of the equipment.

v

 

DO NOT

 use if the equipment is empty

v

 

DO NOT

 operate the tap while the wax is cold as this may damage the tap.

Symbols Defined

WARNING

EARTH

BIOHAZARD

RISK OF

ELECTRIC SHOCK

HOT

SURFACE

Содержание MH8524

Страница 1: ...Cole Parmer WD 200 Series Paraffin Wax Dispenser Instruction Manual ET0003 Version 1 1 ...

Страница 2: ... following conditions v For indoor use only v Use in a well ventilated area v Ambient temperature range 5 C to 40 C 41 F to 104 F v Altitude to 2000 m 6500 ft v Relative humidity not exceeding 80 temperature 31ºC decreasing to 50 temperature 40ºC v Mains supply fluctuations not exceeding 10 of nominal v Overvoltage category II IEC60364 4 443 v Pollution degree 2 IEC664 v Use with a minimum distanc...

Страница 3: ...RDS OR UNSAFE PRACTICES ARE DANGEROUS as they can cause severe personal injury fire or death v DO NOT use combustible substances near hot objects v DO NOT use the equipment in hazardous atmospheres v DO NOT immerse the equipment in water v DO NOT operate or handle any part of the product with wet hands or use on surfaces that may become flooded v NEVER move the product while still connected to the...

Страница 4: ...replace the plug with a suitable alternative THIS OPERATION SHOULD ONLY BE UNDERTAKEN BY A QUALIFIED ELECTRICIAN NOTE Refer to the equipment rating plate to ensure that the plug and fusing are suitable for the voltage and wattage stated The wires in the mains cable are as follows 230V 120V BROWN HOT LIVE BLACK HOT LIVE BLUE NEUTRAL WHITE NEUTRAL GREEN YELLOW EARTH GREEN EARTH The appropriate power...

Страница 5: ...tion 1 Ensure that the power cable is pushed fully into the power supply socket of the wax dispenser 2 Connect the mains plug to the electrical supply 3 Pour the desired quantity of wax into the internal tank The tank has a maximum capacity of 7 5L which is large enough to melt 6 0kg of pelletized histology wax 4 Turn ON the wax dispenser 5 Set the desired temperature Users are recommended to set ...

Страница 6: ...in as quick a time as possible When set at 65o C the instrument will take roughly 90 minutes to fully melt 5kg of pelletized wax with significant quantities of molten wax available for use after as little as 30 minutes 9 Wax is delivered by pulling the tap lever forward The tap has three positions i Off ii On demand pull dispense iii Continuous flow tap locked open CAUTION Do not operate the tap w...

Страница 7: ...d for more thorough cleaning using solvents If removing the filter it is recommended that protective gloves are worn NOTE DO NOT use scouring pads de scaling agents or solvents to clean any parts of this equipment Preventative maintenance should include keeping the product clean by protecting it from spillage contamination or corrosive environments If in doubt please confirm that any intended meth...

Страница 8: ...ontamination certificate Copies of the Certificate are available from the Distributor Manufacturer At the end of its service life the product must be accompanied by a Decontamination Certificate Repairs and Support Any repairs or replacement of parts MUST be undertaken by suitably qualified personnel Only spare parts supplied or specified by Antylia Scientific or its agents should be used Fitting ...

Страница 9: ...with Max 6kg of wax Temperature display range Ambient to 70ºC 1ºC at 20ºC ambient Display resolution 1ºC Accuracy 1 C Ceiling temperature setting 70 C Display Digital Display Control Sensor NTC Thermistor Electrical specification Total power consumption 480W Total power output 5A Temperature control On Off Oven operating temperature range Ambient to 70ºC 1ºC at 20ºC ambient Electricity supply 110 ...

Страница 10: ...10 158 mm 455 mm 455 mm 181mm 181mm 420 5mm 79mm 455 mm ...

Страница 11: ...so we cannot guarantee that interference will not occur in practise Where there is a possibility that injury damage or loss might occur if equipment malfunctions due to radio frequency interference or for general advise before use contact the manufacturer EU Representative address Antylia Scientific GmbH Futtererstraße 16 97877 Wertheim Deutschland Tel 49 9377 9203 0 Email sales coleparmer de UK R...

Страница 12: ...antylia com W coleparmer fr Italy T 39 0 284349215 E it sales antylia com W coleparmer it India T 9122 61394444 E info coleparmer in W coleparmer in China T 1 847 549 7600 E sales antylia com W coleparmer com USA T 1 847 549 7600 E sales antylia com W coleparmer com Canada T 514 355 6100 E info antylia ca W coleparmer ca Other T 1 847 549 7600 Ordering Information Order No Series Model Legacy SKU ...

Отзывы: