This will take you to the setpoint screen. The user has an option to adjust
the minimum and maximum setpoints for current and RPM. The
setpoints being edited will be blinking on the slope. To change which
setpoint to edit use the UP or DOWN to select screen. By default the
minimum (MIN) RPM is set to 0.0 and the current is set to 4.1 mA.
The maximum is set to maximum (MAX) RPM and the current is set to
20.0 mA. The scaling can be inverted, if necessary. The current is linear
between these two points. After both setpoints have been adjusted, save
both setpoints from the upper setpoint setting screen by highlighting the
save icon and pressing ENTER.
To confirm CURRENT OUTPUT MODE is ON, press the BACK key
until the CONTINUOUS PUMP MODE screen appears. The Current
output icon will be displayed.
NOTE:
When Current Output is selected the drive will not start until the
CURRENT OUTPUT is ON and is in CONTINUOUS PUMP
MODE.
NOTE:
The Current Output indicates the Running Command Speed. Use
the Motor Running contacts (normally open) to indicate if pump is
running.
Remote Control
Current Output Menu
(continued)
3-20
MASTERFLEX
®
Digital Pump Drive Operating Manuall
Cole-Parmer
Section 3
Operation