10. Maintenance and cleaning 

Always place the dust cover over your MSS-200 Series microscope after use. Always keep the eyepieces mounted on 
the microscope to avoid dust entering the instrument

10.1 Cleaning the optics 

When the eyepiece lenses are dirty, they can be cleaned by wiping a piece of lens paper over the surface (circular 
movements). If this does not help put a drop of alcohol on the lens paper and wipe it. Never put Isopropyl or alcohol 
directly on the lens! Cole-Parmer has various optical cleaning wipes available

It is not necessary - and not recommended - to clean the lens surfaces at the inner side of the objectives. Sometimes 
dust can be removed with high pressured air

 Caution

 Cleaning cloths containing plastic fibers may damage the coating of the lenses! 

10.2 Maintenance of the stand 

•  

Dust can be removed with a brush. In case the stand or base plate is really dirty, then clean the surface with a 
non-aggressive cleaning product

•  

All moving parts - like the height adjustment or the coaxial coarse and fine adjustment - contain ball bearings 
that are not dust sensitive. With a drop of sewing-machine oil the bearing can be lubricated

10.3 Replacing the fuse 

To change the fuse, follow the procedures below:

•  

Remove the power cord from the back of the microscope

•  

Locate the fuse compartment, which has a fuse image. It is typically 
located close to the power connector

•  

Remove the fuse compartment. To do so, insert a flat head 
screwdriver behind the fuse compartment and gently pry it loose 

•  

Insert the new fuse into the compartment, and then replace the 
fuse compartment back to where it was originally

•  

Power up the microscope and test

•  

Fuse specification: 250 V F1A glass, 5 x 20 mm

Note:

   The fuse may blow in order to prevent internal damage to the microscope and in most cases, replacing the 

fuse with a correctly rated fuse will resolve the issue. However, should you encounter a blown fuse frequently, please 
contact your distributor for further assistance

11. Accessories and spare parts

For current accessories and spares, please check our website coleparmer.com

12. Warranty

These microscopes come with a five (5) year factory warranty against manufacturing defects, covering labor and 
parts. Warranty does not apply for consumable accessories and parts such as bulbs, batteries, fuses, cords, optical 
components, or any add-on accessories such as mechanical specimen holders that are not built into the microscope 
stage as an integral part of the original manufacture, etc. Warranty does not cover microscopes, lenses cameras or 
other accessories that have become inoperable due to dirt or damage due to misuse or lack of maintenance. 

Note:

 Buyers are responsible for return shipping and handling cost for warranty services. Warranty covers parts and 

labor only.

11

Содержание 78904-66

Страница 1: ...Installation and Operating Instructions MSS 200 Series Trinocular Stereozoom Microscopes Models 78904 66 69 78904 66 78904 69...

Страница 2: ...ctives________________________________________________________________________________ 7 4 4 Object stage ______________________________________________________________________________ 7 4 5 Focusing_...

Страница 3: ...with a cover glass never comes in direct contact with the microscope parts In that case prevention of contamination lies in the handling of the slides as long as the slides are decontaminated before...

Страница 4: ...based disinfection products for laboratories Before returning a microscope for repair or maintenance through a Cole Parmer dealer an RMA return authorization form together with a decontamination state...

Страница 5: ...when items are placed on the microscope Keep flammable fluids fabric etc well out of the way Disconnect power Always disconnect your microscope from power before doing any maintenance cleaning assemb...

Страница 6: ...smitted illumination G Object plate with transmitted illumination below it H Stand base I Handle J Photo port K Microscope head L Stand rack pinion version M Focusing knobs only one is visible N Incid...

Страница 7: ...n stands and pillar stands are equipped with an object stage with a semi transparent stage plate G and two object clips N 4 5 Focusing With the two coarse focusing knobs M the object can be sharply fo...

Страница 8: ...ching the tubes The eyepieces can be secured with a screw The exact location of the hexagon screw may differ Each eyepiece has a rubber eyeshade This prevents damage to the lens and stray light The ey...

Страница 9: ...t knob to the lowest magnification Rotate the focusing knob to bring the specimen into focus Turn the zoom adjustment knob to the highest magnification Rotate the focusing knob to bring the specimen i...

Страница 10: ...w through the eyepieces Adjustment can be made by raising lowering the height of the camera C Take an easy to view specimen and focus the image through the microscope s eyepieces with dioptre adjustme...

Страница 11: ...the fuse compartment which has a fuse image It is typically located close to the power connector Remove the fuse compartment To do so insert a flat head screwdriver behind the fuse compartment and gen...

Страница 12: ..._________________________________________________________________ ___________________________________________________________________________________________________ __________________________________...

Отзывы: